The measurement method of gamma ray air absorbed dose rate based on LaBr3(Ce) detector |
CAO Qinjian1, XIA Sanqiang1, HE Meng2, LLU Liye1, XIAO Yunshi1, LI Xiaodun1, JIAO Yan1 |
1. China Institution for Radiation Protection, Taiyuan 030006 China; 2. China Nuclear Power Engineering Co.,Ltd,Beijing 100840 China |
|
|
Abstract Objective Based on the lanthanum bromide scintillator detector, the calculation method of G(E) function was developed to measure the air absorbed dose rate.Methods Firstly, the gamma energy spectrumof the lanthanum bromide detector were simulated and the response functions with different energies were determined with Monte Carlo simulation method. Then, the G(E) function was calculated by the least square method. Finally, the air absorbed dose rate measured by the lanthanum bromide detector using G(E) function conversion method was compared with the theoretical calculation value based on the point source experiments.Results The experimental verification results showed that the relative deviation between thecalculated value with G(E) function and the theoretical calculation value wasmostly controlled within ±6%, which verified the accuracy of G(E) function.Conclusion The results showed that the method of G(E) function could be applied to calculate the gamma radiation dose rate based on the in-situ the gamma spectrum with LaBr3 detector.
|
Received: 15 June 2020
|
|
|
|
|
[1] Moriuchi S, Miyanaga I. A spectrometric method for measurement of low-level gamma exposure dose[J]. Health Phys, 1966, 12(4): 541-551. DOI: 10.1097/00004032-196604000-00009 [2] Terada H, Sakai EJ, Katagiri M. Environmental gamma-ray exposure rates measured by in situ Ge(Li) spectrometer[J]. J Nucl Sci Technol, 1980, 17(4): 281-290. DOI: 10.1080/18811248.1980.9732580 [3] Jae-shik Jun. Calculation of the spectrum-dose Conversion Operator for a HPGEDetector by Means of a Monte Carlo Simulation[J]. J Korean Soc, 1992, 25: 504-509 [4] Park K, Kim J, Lim KT, et al. Improvement of a spectrum-to-dose conversion function for electronic personal dosimeters[J]. J Inst, 2020, 15(2): C02018. DOI: 10.1088/1748-0221/15/02/c02018 [5] 张驰, 王玉东, 周荣, 等. G(E)法与Gravel法处理能谱-剂量转换效果研究[J]. 核电子学与探测技术,2017,37(3):268-273. DOI: 10.3969/j.issn.0258-0934.2017.03.009 Zhang C, Wang YD, Zhou R, et al. The research of G(E)function and gravel algorithm managing spectra-dose conversion[J]. Nucl Electron Detect Technol, 2017, 37(3): 268-273. DOI: 10.3969/j.issn.0258-0934.2017.03.009 [6] 黄金峰, 王莹, 熊文俊, 等. 基于Monte-Carlo模拟和峰形拟合的CdZnTe探测器G(E)函数简便计算方法研究[J]. 原子能科学技术,2017,51(1):165-169. DOI: 10.7538/yzk.2017.51.01.0165 Huang JF, Wang Y, Xiong WJ, et al. Study on simple G(E) function calculation method of CdZnTe detector based on Monte-Carlo method and peak shape fitting[J]. At Energy Sci Technol, 2017, 51(1): 165-169. DOI: 10.7538/yzk.2017.51.01.0165 [7] 王晨毅. 基于G(E)函数法的γ能谱-剂量转换研究应用及计算软件开发[D]. 衡阳: 南华大学, 2018. Wang CY. Gamma-ray spectrum-dose conversion research and computation software compilation based on G(E) function method[D]. Hengyang: University of South China, 2018. [8] 邓中华, 王媛, 赵永生, 等. MC法创建(LaBr3:Ce3+)探测器能谱-剂量转换G(E)函数[J]. 核电子学与探测技术,2018,38(3):441-444. DOI: 10.3969/j.issn.0258-0934.2018.03.027 Deng ZH, Wang Y, Zhao YS, et al. Study on G(E) function of the (LaBr3:Ce3+) detector by Monte Carlo method[J]. Nucl Electron Detect Technol, 2018, 38(3): 441-444. DOI: 10.3969/j.issn.0258-0934.2018.03.027 [9] 贾明雁, 程建平, 冯天成. 就地γ谱仪测量原理及应用[M]. 北京: 中国原子能出版社, 2013. Jia MY, Cheng JP, Feng TC, et al. Principle and Application of in-situ Gamma Spectrometer Measurement[M]. Beijing: China Atomic Energy Press, 2013. [10] 曹勤剑, 夏三强, 刘立业, 等. 溴化镧探测器的晶体表征与全能峰效率计算[J]. 辐射防护,2016,36(6):364-367, 416 Cao QJ, Xia SQ, Liu LY, et al. Size adjustment and calculation of full-energy peak efficiency for the LaBr3 scintillation detector[J]. Radiat Prot, 2016, 36(6): 364-367, 416
|
|
|
|