Measurement and Analysis of Leakage Radiation of 18 Medical Electron Accelerators in M area 24 Points |
XIA Yuan-fu1, ZHU Jian-guo2, YANG Shu-hui2, ZHANG Lin2, GAO Ze-yu2, LU Feng2, NIU Fei2 |
1. Linyi Cancer Hospital, Linyi 276000 China; 2. Institute of Radiation Medicine, SDAMS |
|
|
Abstract Objective To analyse and evaluate the protective performance of medical electron accelerator through measuring and analyzing the leakage radiation levels of some different types of medical electron accelerators in M area 24 points.Methods The ratios of X-ray leakage radiation of 18 medical electron accelerators were tested and studied using thermoluminescent dosemeter (LiF(Mg,Cu,P) plastic tube) according to the measure method of GBZ 126-2011 strictly.Results About the ratios of X-ray leakage radiation in M 24 points area of 18 medical electron accelerators,the range of maximums is from 0.007% to 0.073% (the requirement of GBZ126-2011 is ≤ 2%)and the range of average is from 0.005% to 0.059%(the requirement of GBZ126-2011 is ≤ 0.75%),except to the seventh accelerator for which the beam limiter is damaged.Conclusion The maximum and the average ratio of X-ray leakage radiation of 17 medical electron accelerators meet the relevant national occupational standards, but with the exception of 17.
|
Received: 04 May 2017
|
|
|
|
|
[1] 马永忠,王时进,苏旭,等.某医院18 MV医用加速器靶外辐射剂量水平及其影响因素[J].中国职业医学,2007,34(2);159-160. [2] 中华人民共和国卫生部,GBZ 126-2011电子加速器放射治疗放射防护要求[S].北京:中国标准出版社,2011. [3] IAEA. Absorbed dose determination in photon and electron beams an international code of practice[R].Technical reports series No.277, sec. ed. International Atomic Energy Agency, Vienna, 1997. [4] 赵新春,周艳,张会敏,等.广西医用电子加速器验收检测泄漏辐射水平分析[J].中国医学装备,2014,2(2):9-12. [5] 杜翔,曹兴江,杨春勇,等.剂量分布仪对医用电子直线加速器X射线泄漏辐射检测方法的探讨[J].中国医学装备,2014, 11(7):4-6. [6] Adnan K, Jaradat, Peter J, et al.Measurement of the leakage radiation from linear accelerators in the backward direction from 4,6,10,15 and 18MV X-Ray energies[J].Health Physica,2007,92(4):387-395. |
|
|
|