Analysis of quality control status of medical X-ray diagnostic equipment in a military region |
KOU Qinghe1, XIE Huaijiang1, TONG Yushe1, MA Li2 |
1. Sheng Yang Military Region Center for Disease Control and Prevention, Shenyang 110034 China; 2. Epidemic Prevention Teams of the Support Department in Morthern Theater Command of the LAAF |
|
|
Abstract Objective To acquaint the basic status of quality control of X-ray diagnostic equipment in a military region so as to improve radiation protection level.Methods According to the related national criterions, quality control tests for medical X-ray diagnostic equipment were carried out in the medical institutions in a military region.Results The qualified rates of radiography and fluoroscopy equipment were 84.7%,92.4% respectively. The total qualified rate was 88.7%. There was no statistical difference among different levels of medical institutions.Conclusion There existed some problems in quality control of X-ray diagnostic equipment in the military region. It is necessary to strengthen the quality control tests of X-ray diagnostic equipment, especially for the low level institutions.
|
Received: 24 November 2017
|
|
|
|
|
[1] 中华人民共和国卫生部.WS 76-2011医用常规X射线诊断设备影像质量控制检测规范[S].北京:中国标准出版社,2011. [2] 郑钧正.放射诊疗的防护与安全是医疗设备应用安全和质量管理的重点[J].中国医疗设备,2010,25(9):3-5. [3] 王俊生,张怡,赵坤,等.深圳市常规医用X射线诊断设备影像质量状况分析[J].职业卫生与应急救援,2013,31(3):129-131. [4] 郑森兴,郭进瑞,张燕,等.2012-2013年福建省医用X射线诊断设备质量控制检测结果分析[J].中国辐射卫生,2014,23(6):522-523. [5] 刘洪博,姜潮,崔滨. 2013年哈尔滨市放射诊断设备性能检测与评价[J].中国公共卫生管理,2015,31(5):753-754. [6] 王欢,王新明,朱维杰,等.2015年北京市X射线诊断设备质量控制现状研究[J].首都公共卫生,2016,10(4):151-154. |
|
|
|