Performance quality control detection and result analysis of SPECT equipment |
HU Bo, LIAO Xiangdong, LI Changhong, LIU Ran, MA Qiao |
Sichuan Provincial Center for Disease Control and Prevention, Chengdu 610041 China |
|
|
Abstract Objective To understand the performance status of SPECT and improve the quality control of the SPECT by testing 9 SPECT equipments in Sichuan province.Methods According to the relevant national and NEMA standards, using SPECT performance detection mode and image analysis code, the performance of the 9 SPECTs were measured including their tomographic spatial resolution, system planar sensitivity, system spatial resolution, intrinsic counting rate performance, intrinsic spatial resolution, intrinsic spatial resolution, intrinsic uniformity, intrinsic spatial differential linear.Results In the 7 performance indicators, the intrinsic uniformity and the intrinsic spatial differential linear qualification rate are relatively low, the pass rate is only 44.4% and 33.3%, intrinsic counting rate performance and the intrinsic spatial resolution pass rate are 88.9%, and the other indexes pass rate are 100%.Conclusion Users should have some daily maintenance skills and awareness, take quality control measures for equipment, emphasize the importance of performance testing for SPECT devices, so as to improve the quality of clinical diagnosis.
|
Received: 12 March 2018
|
|
|
|
|
[1] 国家质量监督检验检疫总局.GB/T 18989-2013放射性核素成像设备性能和试验规则伽马照相机[S].北京:中国标准出版社,2013. [2] 国家质量监督检验检疫总局.GB/T 18988.2-2013放射性核素成像设备性能和试验规则第2部分:单光子发射计算机断层装置[S].北京:中国标准出版社,2013. [3] NEMA.Standards publication NU1-2007:performance measurements of gamma cameras[S].NEMA,2007. [4] 齐雪松,江泓,王式琦, 等.核医学影像设备的质控方法[J].中华放射医学与防护杂志,2003,(2):123-125. [5] 谭展,马睿,谭光享,等.广东省21台SPECT机性能检测结果分析[J].中国辐射卫生,2017,26(06):665-667. [6] 林祥通.双探头SPECT、FDG SPECT仪质量控制中应注意的几个问题[J]. 中华核医学与分子影像杂志,2001,21(1):6. [7] 高泽宇,朱建国,杨淑慧,等.SPECT设备固有均匀性性能测试结果分析[J].中国辐射卫生,2017,26(04):426-428. [8] 卢峰,宋钢,李全太,等.SPECT设备性能测试及结果分析[J].中国辐射卫生,2014,(1):26-28. |
|
|
|