Investigation on the Extremity Dose to Interventional Staff by Using Interventional Devices with Single and Double X-ray Tubes |
HUANG Zhuo, FAN Yao-hua, LI Wen-yan, OU Xiang-ming, YUE Bao-rong, XU Hui |
Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088 China |
|
|
Abstract Objective To establish the method for extremity dose measuring for interventional staff, and compare the extremity dose to interventional staff on traditional interventional devices with single X-ray tube to that on the new devices with double tubes, and thus support the monitoring of the doses and the improvement of the radiological protection.Methods Thethermoluminescence dosimeters (TLD) were calibrated with personal dose equivalent Hp(0.07). The extremity doses of 231 person-time from 4 kinds of interventional procedures were measured on 2 interventional devices with single X-ray tube and 3 devices with double X-ray tubes. The doses obtained on different kinds of interventional devices were analyzed.Results The TLD shows good linear fitting with Hp(0.07) calibration.The extremity doses, which in the therapeutic procedures were higher than those in the graphic ones, perform positively relevance to the exposure time.Comparing the same kind of interventional procedures on devices with double and single X-ray tubes, the former shows shorter exposure time with higher extremity dose to interventional staff than the latter.Conclusion The interventional devices with double X-ray tubes significantly improve the efficiency of the interventional procedures while increase the dose to interventional staff. The radiological protection should be strengthened to protect the health of the interventional staff.
|
Received: 06 April 2017
|
|
|
|
|
[1] Chodick G, Bekiroglu N, Hauptmann M, et al. Risk of cataract after exposure to low doses of ionizing radiation:a 20-year prospective cohort study among US radiologic technologists[J]. Am J Epidemiol, 2008, 168(6):620-631. [2] Stewart FA, Akleyev AV, Hauer-Jensen M, et al. ICRP Publication 118. ICRP statement on tissue reactions and early and late effects of radiation in normal tissues and organs-threshold doses fortissue reactions in a radiation protection context[J]. Ann ICRP, 2012,41(1-2):1-322. [3] 赵红枫,岳保荣,尉可道.介入放射学程序中职业人员手部受照剂量研究状况[J].中华放射医学与防护杂志,2013, 33(2), 218-220. [4] Szumska A, Budzanowski M, Kopec' R. Occupational exposure to the whole body, extremities and to the eye lens in interventional radiology in Poland, as based on personnel dosimetry records at IFJ PAN[J]. RadiatPhysChem, 2014, 104(1):72-75. [5] 于夕荣,柳怡,张巍,等.介入放射诊疗操作人员个人剂量水平调查[J].中国医学装备, 2008, 5(10):23-25. [6] Struelens L, Schoonjans W, Schils F, et al. Extremity and eye lens dosimetry for medical staff performing vertebroplasty and kyphoplasty procedures[J]. J Radiol Prot, 2013, 33(3):635-645. [7] Domienik J, Brodecki M, Carinou E, et al. Extremity and eye lens doses in interventional radiology and cardiology procedures:first results of the ORAMED project[J]. RadiatProtDosim, 2011, 144(1-4):442-447. [8] 赵智慧,周开建,张京战,等.介入诊疗中放射工作人员受照剂量调查[J].职业与健康, 2012, 28(22):2731-2732. [9] 赵军,李秋香,李斌,等.经桡动脉和股动脉路入途径行冠状动脉造影时术者手部受照剂量的对比[J].中华放射医学与防护杂志, 2011, 31(4):495-496. [10] International Commission on Radiological Protection.The 2007 Recommendations of the International Commission on RadiologicalProtection[R]. ICRP Publication 103. Oxford:PergamonPress,2007. [11] Whitby M, Martin CJ. A study of the distribution of dose across the hands of interventional radiologists and cardiologists[J]. Br J Radiol, 2005, 78(927):219-229. [12] Koukorava V, Carinou E, Simantirakis G, et al. Doses to operators during interventional radiology procedures:focus on eye lens and extremity dosimetry[J]. RadiaProtDosim, 2011, 144(1-4):482-486. [13] 欧向明,范瑶华.腕式剂量计校准的标准辐射场的建立及其应用[J].中华放射医学与防护杂志,2016, 36(9):703-706. [14] 中华人民共和国国家卫生和计划生育委员会.GBZ 130-2013医用X射线诊断放射防护要求[S].北京:人民卫生出版社,2013. |
|
|
|