Comparative study on the entrance surface dose of oral COPT and CBCT subjects |
TIAN Qingxiang1,2, LOU Yun1,2, WANG Hongfang1,2, WANG Xinming1,2 |
1. Beijing Central for Preventive Mediciine Research, Beijing 100013 China; 2. Beijing Central for Disease Control and Prevention |
|
|
Abstract Objective For comparison of the entrance surface doses in the two types of computed orthopantomography(COPT) and cone beam CT(CBCT).Methods The thermoluminescence dosimeter (TLD) was used to measure the entrance surface doses (ESD) undergoing oral scans with COPT and CBCT in two types of oral CT. In the experiments, the TLD elements were placed at 12 focused points of the head phantom, and their doses were measured by the RGD-3B® reading system.Results For the two types of CT under COPT routinely using conditions, the ESDs of CBCT were all higher than those of COPT, and the ratios ranged from 1.2 to 9.8 and 1.8 to 14.4 for type A and type B device, respectively.Conclusion Dentists should realize different doses from the two inspection methods of COPT and CBCT, and select appropriate exposure conditions, guide the examinees to use the personal protective equipment correctly, in order to reduce radiation hazard to the lowest possible level.
|
Received: 16 June 2017
|
|
|
|
|
[1] 欧向明,赵士安,李明生,等.X射线诊断入射体表剂量监测用热释光剂量计的质量控制[J].中国医学装备,2012,9(5):4-7. [2] 武大鹏,王宏芳,朱维杰,等.北京地区X射线摄影受检者体表入射剂量调查分析[J].中国辐射卫生,2014,23(1):17-19. [3] 朱志贤,韩发明,唐文祥,等.牙科X射线全景摄影所致受检者体表剂量测量方法研究.中华放射医学与防护杂志,2000,20(增刊):75-77. [4] Horowitz Y,Olko P. The effects of ionization density on the thermoluminescence response(efficiency) of Lif:Mg,Ti and Lif:Mg,Cu,P.Radiat Prot Dosimetry,2004,109(4):331-348. [5] 李刚.口腔颌面锥形束CT放射剂量与放射防护[J].中国实用口腔科杂志,2011,4(10):577-580. [6] 曲兴民,李刚,JB Ludlow,等.ProMax 3D口腔颌面锥形束CT不同扫描参数设置下的辐射剂量研究[J].中华口腔医学杂志,2011,46(9):562-563. |
|
|
|