Analysis of CT machine quality control and radiation protection test results in Luoyang from 2016 to 2017 |
WANG Xinai1, WANG Haisheng1, LI Ruzhen2, LI Zhe1 |
1. Luoyang Center for Disease Control and Prevention, Luoyang 471000 China; 2. Luoyang Municipal Bureau of Health and Family Planning Supervision |
|
|
Abstract Objective Understand testing index of CT machine quality control and protection status of CT machine room in Luoyang city.Methods By random sampling method, 55 CT machines in 45 hospitals in Luoyang in 2016, 37 CT machines in 32 hospitals in Luoyang in 2017 were tested for quality control, radiation level around the machine room was detected, and the test results were analyzed and evaluated.Results The test results of quality control showed that the qualified rate in 2016 was 63.60% and that in 2017 was 89.20%, with statistically significant differences (P < 0.05). In 2016, the qualified rate of primary, secondary and tertiary hospitals was between 60.00% and 65.52%; in 2017, the qualified rate of primary hospitals was 78.90%; and the qualified rate of both secondary and tertiary hospitals was 100.00%. In 2016, the qualified rate of detection by a 16-row less CT machine was 51.85%, 70.60% by 16-row CT machine, and 81.80% by a 64-row CT machine or above. In 2017, the qualified rate of detection was 86.60%, 85.70% and 100.00% respectively. The qualification rate of the single index of the quality control test of the CT machine was 76.36% in 2016, while the rest were above 90.0%. The qualification rate of the single index in 2017 was all above 90.0%. According to the radiation level test results around the CT machine room, the qualified rate in 2016 was 98.04%, and the qualified rate in 2017 was 97.30%.Conclusion In 2016, the qualified rate of CT machine quality control in luoyang was relatively low. Although it was greatly improved in 2017, the supervision still needs to strengthen, with the emphasis on the supervision of primary and lower level hospitals.
|
Received: 20 July 2019
|
|
|
|
|
[1] 赵进沛, 杨会锁. 放射诊疗卫生防护及其监督监测[M]. 北京: 军事医学科学出版社, 2013. [2] 余子丽, 周正东, 张雯雯, 等. 多能光子计数X-CT能量加权图像重建方法[J]. 中国医学物理学杂志,2016,33(2):141-145 [3] 中华人民共和国卫生部, 中国国家标准化管理委员会. GB17589—2011X射线计算机断层摄影装置质量保证检测规范[S]. 北京: 中国标准出版社, 2012. [4] 中华人民共和国卫生部. GBZ165—2012X射线计算机断层摄影放射防护要求[S]. 北京: 中国标准出版社, 2013. [5] 中华人民共和国国家卫生和计划生育委员会. GBZ130—2013医用X射线诊断放射防护要求[S]. 北京: 中国标准出版社, 2014. [6] 李宾, 王川健, 朱仔运, 等. 海南省CT机质量控制和放射防护检测结果分析[J]. 中国职业医学,2017,44(5):622-624 [7] 马桥, 刘德明. 四川省CT机质量控制与辐射防护检测结果分析[J]. 职业卫生与病伤,2015,30(3):161-163 [8] 牛振, 张继勉. 天津市CT机使用质量控制检测结果[J]. 职业与健康,2013,29(18):2397-2398 [9] 王新明, 朱维杰, 韩浚, 等. 北京市80台放射诊断设备性能检测与分析[J]. 中国辐射卫生,2014,23(1):28-31 [10] 李士正, 楚彩芳, 李庆新, 等. 南阳市78台CT机质量控制检测与分析[J]. 中国辐射卫生,2018,27(5):476-478 [11] 石建华, 石建忠. 山西省部分医疗机构医用X射线诊断设备影像质量控制检测结果分析[J]. 中国辐射卫生,2018,27(3):231-233 [12] 中华人民共和国国家卫生健康委员会. WS519—2019X射线计算机体层摄影装置质量控制检测规范[S]. 北京: 中国标准出版社, 2019.
|
|
|
|