Analysis of the quality control testing results for CT machines in Xinxiang City in 2016 |
HUO Haiwei, ZHANG Bingjie, MAO Yuxuan, WANG Xuan, WANG Yan, NIE Jing, LIU Tongzhen, JIA Tianhe, WANG Jianwei, DING Weimin |
Xinxiang Institute for Occupational Disease Prevention and Treatment, Xinxiang 453000 China |
|
|
Abstract Objective To analyze the quality control test results of the CT machines in Xinxiang City in 2016, so as to promote the development of radiological health work in the city. Methods The CT machines of the entrusted unit in Xinxiang city were tested for quality control according to GB 17589-2011. Results The qualified rate of all ten testing indexes of the CT machine was 85.5%. Conclusion On the whole,the quality control testing results of the CT machine in Xinxiang City are generally satisfying, but there are certain problems in the quality control status of the second-hand CT machine. The maintenance and update of the second-hand CT machine should be further strengthened, so as to protect the safety and health of the examinee and patient.
|
Received: 02 February 2019
|
|
|
|
|
[1] 李广喜,王胜利.菏泽市198台医用CT机影像质量控制检测结果分析[J].中国卫生标准管理,2019,10(1):124-127. [2] 李士正,楚彩芳,李庆新,等.南阳市78台CT机质量控制检测与分析[J].中国辐射卫生,2018,27(5):476-478. [3] 刘科,胡涵,侯占仙.自贡市X射线CT机质量控制检测与放射防护监测结果分析[J].中国辐射卫生,2018,27(6):552-554. [4] 毛喻萱,王电辉,张冰洁,等.新乡市2013年CT机影像质量控制检测结果分析[J].影像与检验,2015,1(6):200-204. [5] 林海辉,杨宇华,谭展,等.2014年广东省CT机质量保证检测结果分析[J].中国辐射卫生,2017, 26(3):309-315. |
|
|
|