Analysis of 320 CTs' performance test and radiation protection test |
ZHU Weijie1, XU Hui2, HAN Jun3, YU Jun3, CHEN Juan1 |
1. Peking University, School of Public Health, Beijing 100191 China; 2. National Institute for Radiological Protection, China CDC; 3. Beijing Center for Disease Prevention and Control |
|
|
Abstract Objective To detect and analyze the performance and protection quality control of CT equipment in Beijing, and provide reference for improving the performance and protection quality of CT equipment and promoting the reasonable allocation of CT equipment in Beijing.Methods 320 CT equipment in Beijing were detected for performance status and radiation level respectively according rules and standards in GB 17589-2011 "Specifications for quality assurance test for computed tomography X-ray scanners" and GBZ 130-2013 "Requirements for radiological protection in medical X-ray diagnosis". SPSS 20.0 software was used for analysis.Results The testing results of 320 CT equipment's performance and protection quality showed that the qualified rate of performance test and protection test was both 99.7%, and the proportion of ≥ 64 rows CT increased by years.The comparison between imported CT and domestic CT revealed that, the high contrast resolution (high contrast algorithm) of imported CT was better than domestic CT (P=0.03), while the low contrast detectability index of domestic CTwas better than imported CT (P=0.03).Conclusion The qualified rate of CT equipment's performance and protection quality in Beijing was high, which should be credited to the persistently monitoring and constantly updated measures, such as "constancy test" and "networks monitoring". To optimize the CT equipment's allocation and improve the work efficiency, increasing the proportion of domestic CT and ≥ 64 rows CT could be considered.
|
Received: 12 June 2019
|
|
|
|
|
[1] 刘科,胡涵,侯占仙.自贡市X射线CT机质量控制检测与放射防护监测结果分析[J].中国辐射卫生,2018,27(6):552-554. [2] 徐丹丹,袁璐.武汉市放射诊疗机构2015年度CT使用现状调查及分析[J].中国辐射卫生,2017,26(6):674-679. [3] 李士正,楚彩芳,李庆新,等.南阳市78台CT机质量控制检测与分析[J].中国辐射卫生,2018,27(5):476-478. [4] 林海辉,杨宇华,谭展,等.2014年广东省CT机质量保证检测结果分析[J].中国辐射卫生,2017,26(3):309-311,315. [5] 徐辉,岳保荣,尉可道,等.我国CT扫描检查中受检者剂量调查结果与分析[J].中华放射医学与防护杂志,2019,39(3):213-217. [6] 中华人民共和国卫生部. GB 17589-2011 X射线计算机断层摄影装置质量保证检测规范[S].北京:中国标准出版社,2012. [7] 中华人民共和国国家卫生和计划生育委员会. GBZ 130-2013医用X射线诊断放射防护要求[S].北京:中国标准出版社,2014. [8] 朱维杰,王新明,韩浚,等.基于北京市141台CT性能检测结果的差异分析与启示[J].中华放射医学与防护杂志,2018,38(6):439-442,460. [9] 杨秀江.CT空间分辨率和低对比度分辨率的检测及其影响因素[J].科技视界,2018(14):72-73. [10] 李庚,高关心,夏慧琳.CT空间分辨率和低对比度分辨率的检测及其影响因素[J].中国医疗设备,2010,25(1):7-9. [11] 伍健,耿建华.CT设备质量控制的现状与展望[J].中国医学装备,2018,15(11):163-166. [12] 北京市卫生和计划生育委员会.北京市大型医用设备配置与使用管理白皮书[R/OL].2015. http://www.bjchfp.gov.cn/xwzx/wnxw/201512/t20151231_130171.htm. |
|
|
|