The study about the setup error and its effect on dose distribution in IMRT for prostate cancer patients |
GAO Xuan, LI Kexin, JU Yongjian, CAO Liyuan |
Department of Radiation Oncology, The First People's Hospital of Nantong City, Nantong 226001 China |
|
|
Abstract Objective To Analyze the setup error and its effect on dose distribution in IMRT for prostate cancer patients.Methods 45 prostate cancer patients underwent CBCT as the first treatment and compared with the original treatment plan to get the value of bed shifting. Then the setup error were calculated though these values, and patients were divided into two groups, A (≤ 6mm) and B (> 6mm). Then the patients'differences of age, BMI, bladder and rectum volume changes in A and B were compared. Also after bed shifting, the dose parameters of target and OARS were recalculated and compared with that calculated from original treatment plan.Results 1. The bed shifting values in X, Y and Z directions of 45 patients are 0.020 ±0.407 cm, -0.031 ±0.424 cm and -0.107±0.273 cm individually. 2. The difference of BMI between A (23 patients) and B (22 patients) has statistically significant, but the difference of age, bladder and rectum volume changes have no statistically significance. 3. Compared the dose parameters between original plan and latter plan after shifting the treatment bed, the differences of PTV95, the average dose of PTV, CTV95, average dose of CTV and the bladder V60 all have statistically significance, But the differences of rectum V60 and femoral head V20 have no significant difference.Conclusion In prostate cancer patients IMRT,the setup error must be care about, it may be related to the patient's BMI, but has limited/no relationships with the ageand bladder rectal volume changes in this study, and it must be corrected by shifting treatment bed, otherwise it will cause the dose distribution error in the target area and OARs.
|
Received: 05 December 2017
|
|
|
|
|
[1] 邱健健, 张笑颖, 莫凡,等. 新型真空垫在盆腔肿瘤放疗体位固定中的应用评估[J]. 中华放射医学与防护杂志, 2017, Vol.37Issue(11):838-842. [2] 黎杰, 陈川东, 康盛伟,等. 不同分辨率CT图像Sentinel摆位系统误差研究[J]. 中华放射肿瘤学杂志, 2016, 25(8):851-854. [3] 高丽娟, 黄嘉敏, 黄峻,等. Orfit架和真空袋对宫颈癌放疗摆位误差比较[J]. 中华放射肿瘤学杂志, 2017, 26(9):1080-1083. [4] Ariyaratne H, Chesham H, Pettingell J, et al. Image-guided radiotherapy for prostate cancer with cone beam CT:dosimetric effects of imaging frequency and PTV margin[J]. Radiotherapy & Oncology, 2016, 121(1):103-108. [5] Kanakavelu N, Jebaseelan S J. Determination of patient set-up error and optimal treatment margin for intensity modulated radiotherapy using image guidance system[J]. Journal of B.u.on. Official Journal of the Balkan Union of Oncology, 2016, 21(2):505. [6] 李克新, 鞠永健, 王高仁,等. 4种图像配对方式对胸腹部肿瘤放射治疗摆位误差比较研究[J]. 生物医学工程与临床, 2016(2):154-156. [7] 高献书, 亓昕. 前列腺癌根治性放疗靶区勾画共识与争议[J]. 中华放射肿瘤学杂志, 2014, 23(4):279-281. [8] Standards B. Radiotherapy equipment-Coordinates, movements and scales[J]. 2002.(IEC1217). [9] 莫小眯, 肖爱农, 伍海彪,等. 千伏级锥形束kV-CBCT在鼻咽癌调强放疗摆位误差的研究[J]. 中国辐射卫生, 2016, 25(2):206-208. [10] 沙翔燕, 李高峰, 李雪南, 等. 在线校正对前列腺癌图像引导调强放疗剂量的影响[J]. 中华放射医学与防护杂志, 2013, 33(002):163-165. [11] Oh Y K, Baek J G, Kim O B, et al. Assessment of setup uncertainties for various tumor sites when using daily CBCT for more than 2200 VMAT treatments[J]. Journal of Applied Clinical Medical Physics, 2014, 15(2). [12] Rico M, Martínez E, Pellejero S, et al. Influence of different treatment techniques and clinical factors over the intrafraction variation on lung stereotactic body radiotherapy[J]. Clinical & Translational Oncology, 2016, 18(10):1-8. [13] Lin L L, Hertan L, Rengan R, et al. Effect of body mass index on magnitude of setup errors in patients treated with adjuvant radiotherapy for endometrial cancer with daily image guidance.[J]. Int J Radiat Oncol Biol Phys, 2012, 83(2):670-675. [14] Adamczyk M, Piotrowski T, Adamiak E, et al. Dosimetric consequences of prostate-based couch shifts on the precision of dose delivery during simultaneous IMRT irradiation of the prostate, seminal vesicles and pelvic lymph nodes[J]. Physica Medica, 2014, 30(2):228-233. [15] Hirose Y, Nakamura M, Tomita T, et al. Evaluation of different set-up error corrections on dose-volume metrics in prostate IMRT using CBCT images[J]. Journal of radiation research, 2014, 55(5):966-975. [16] Liu G, Zhang S, Ma Y, et al. Effects of error on dose of target region and organs at risk in treating nasopharynx cancer with intensity modulated radiation therapy[J]. Pakistan Journal of Medical Sciences, 2016, 32(1):95-100. |
|
|
|