Comparative study on proton and heavy lon accelerator's dose distribution of radiation field measured with EBT3 film and radiation optical dosimetry system |
YUAN Jilong, CHENG Jinsheng |
Key Laboratory, Institute for Radiological Protection and Nuclear Safety, Chinese Center for Disease Control and Prevention, Beijing 100088 China |
|
|
Abstract Objective Comparative analysis of the quality control measurements of the offline EBT3 film analysis and the online real-time analysis radiation optical dosimetry system in proton and heavy ion accelerator's radiation field dose distribution.Methods For active spot scan protons and carbon ions, the spread-out Bragg peak (SOBP) is 3 cm, and the range is 7 cm, 16 cm, 29 cm, respectively, radiotherapy planning system plans 100 mm×100 mm radiation fields, Using EBT3 film and radiation optical dosimetry to measure the dose distribution of the radiation field under various conditions, the EBT3 are scanned after 24 hours' irradiation and the optical system's image are acquired by real time shooting. The comparative dosimetry index is the Full width at half maximum (FWHM).Results Through comparative analysis, regardless of physical significance or statistical analysis, both detectors could measure the dose distribution of proton and heavy ion radiation field, and the results have good consistency and accuracy.Conclusion Both detectors meet the measurement requirements for proton and heavy ion radiation field's dosimetry parameters, Radiation optical dosimetry is capable of on-line dose distribution of radiation field measurement in real time, and more suitable for the development of the quality control of radiation therapy.
|
Received: 19 March 2018
|
|
|
|
|
[1] Giordanengo S, Garella M A, Marchetto F, et al. The CNAO dose delivery system for modulated scanning ion beam radiotherapy[J]. Medical Physics, 2015, 42(1):263-275. [2] 俞顺飞,程金生,李开宝,等.胶片剂量计[J].中国辐射卫生,2008,(2):244-245. [3] 卢峰,宋钢,李全太,等.27台医用电子加速器辐射野半影的胶片法检测结果与分析[J].中国辐射卫生,2014,(2):97-99. [4] Sorriaux J, Kacperek A, Rossomme S, et al. Evaluation of Gafchromic EBT3 films characteristics in therapy photon, electron and proton beams.[J]. Phys Med. 2013, 29(6):599-606. [5] Krzempek D, Mianowska G, Bassler N, et al. Calibration of Gafchromic EBT3 film for dosimetry of scanning proton pencil beam (PBS)[J]. Radiation Protection Dosimetry, 2018,1:1-5. [6] Russo S, Mirandola A, Molinelli S, et al. Characterization of a commercial scintillation detector for 2-D dosimetry in scanned proton and carbon ion beams.[J]. Physica Medica, 2017, 34:48-54. [7] 杨锋, 张湘, Wen Chien Hsi,等. 扫描式质子束成像仪研制及其光学特性研究[C]//中国医学装备协会学术年会. 2016. [8] 袁继龙, 李明生, 程金生. 利用EBT3胶片测量主动式点扫描质子重离子放疗设备照射野一致性研究[J]. 中华放射医学与防护杂志, 2016, 36(8):626-629. [9] Mori Y, Isobe T, Yamaguchi Y, et al. Development of simple high-precision two-dimensional dose-distribution measurement method for proton beam therapy using imaging plate and EBT3[J]. Australasian Physical & Engineering Sciences in Medicine, 2016, 39(3):1-10. [10] 宋文杰,王桂玲,马惠芳, 等.200kV重离子加速器加速氩时辐射场测量与安全评价[J].中国辐射卫生,2001,(2):75-76. [11] Nichiporov D, Hsi W, Farr J. Beam characteristics in two different proton uniform scanning systems:a side-by-side comparison[J]. Medical Physics, 2012, 39(5):2559-2568. [12] Fonseca G P, Podesta M, Bellezzo M, et al. Online pretreatment verification of high-dose rate brachytherapy using an imaging panel[J]. Physics in Medicine & Biology, 2017, 62(13):5440-5461. |
|
|
|