Study on the Quality Control and Radiation Protection for Medical Diagnostic X-ray Machines in Nanchang |
CHENG Wen-juan, GUO Ling-ling, FENG Xiao-wu, JIN Li |
Nanchang Center for Disease Control and Prevention, Nanchang 330038 China |
|
|
Abstract Objective To investigate and estimate the status of the quality control and radiation protection for medical diagnostic X-ray machines in Nanchang.Methods Quality control inspecting and on-spot radiation protection monitoring were conducted by relative national standard methods.Results The total qualified rate of 50 X-ray machines was 66.0%. The qualified rate of domestic and imported machines was 59.4% and 84.6%. And the qualified rate of city-, county-, township-, private- and the staff hospitals were 77.8%, 85.7%, 0.0%, 64.3% and 73.3% respectively. There were eight main quality control indicators included in this study in which the edges deviation and centres deviation of lighting filed to radiation filed had serious problem. The radiation level of one machine room door point from a township-level hospital in total 324 points exceeded the national standards limit.Conclusion We should strengthen the management of X-ray machines to solve the problem about quality control.
|
Received: 11 August 2016
|
|
|
|
|
[1] 中华人民共和国卫生部.放射诊疗管理规定[S].北京,2006-02-28. [2] 国家质量监督检验检疫总局.GB 18871-2002. 电离辐射防护与辐射源安全基本标准[S].北京:中国标准出版社,2002. [3] 中华人民共和国卫生部.WS 76-2011 医用常规X射线诊断设备影像质量控制检测规范[S].北京:中国标准出版社, 2011. [4] 中华人民共和国卫生部.GBZ 130-2013 医用X射线诊断放射防护要求[S].北京:中国标准出版社 2013. [5] 李建军,杨杰,张冰洁等.387台医用诊断X射线机影像质量控制调查分析[J].中国辐射卫生,2014,23(6):534-536. [6] 刘福光,黄伯越, 王丛蕾,等.肇庆市医用X射线诊断装置质量控制检测结果与分析[J].中国辐射卫生,2016,25(3):324-326. [7] 陈以水,王琦,李南翔,等.江西省医用诊断X射线机性能现状调查分析[J].中华放射医学与防护杂志,2011,31(6):702-703. [8] 王心韬,施文华.福州市医用X射线机影像质量控制检测结果与分析[J].中国辐射卫生,2015,24(6):626-628. |
|
|
|