Investigation and analysis of the Dose of 160 Cases of Digital Mammography in Guangxi |
LIANG Ting, XIE Ping, CHEN Zhang-fan, CHEN Fa-xiang, DONG Ying, WU Ying-yu |
Guangxi Center for Disease Prevention and Control, Naning 530028 China |
|
|
Abstract Objective To investigate the dose level of direct digital mammography in Guangxi, explore the factors that influence the image quality and dose in direct digital mammography.Methods A survey was carried out on 160 cases of mammography exposure recorded from 4 medical facilities in Guangxi. The personal information, photography posture, breast compression thickness, exposure parameters and the average breast dose (AGD) were collected. ANOVA and non-parametric test were used to analyse the data.Results The mean value of breast compression thickness in the collected 160 mammography parameters was 42 mm, median was 43 mm,the mean value of AGD was 2.0 mGy,median was 1.5 mGy. Significant difference exists in difference breast compression thickness of AGD (F=20.036,P<0.05). Breast compression thickness increases, the higher of AGD. The mean value AGD of MLO was 1.99 mGy, and the mean value AGD of CC was 2.0 mGy.No significant difference in the AGD data was found between the craniocandal and mediolateral oblique (F=0.685,P<0.05).Conclusion The value of AGD in Direct Digital Mammography of Guangxi was hight. Breast compression thickness,different anode/filter combinations and photography conditions have effect on AGD. For the consideration of the abore factors, the dose of the subjects shonld be controlled after the lowest possible leved.
|
Received: 22 May 2017
|
|
|
|
|
[1] 杜翔,王进.江苏省乳腺X射线摄影医疗照射剂量调查[J].环境与职业医学,2014,31(002):109-111. [2] 刘荫华,徐玲,叶晶明,等.加快建立符合我国卫生经济学的乳腺癌早诊体系[J].中国实用外科杂志,2009,3:220-222. [3] 王艳平,吕锐,王海阔,等.低剂量乳腺X线摄影图像质量和辐射剂量的对比研究[J].中国病案, 2015(11):84-87. [4] 柳杰,刘佩芳,王红彬,等.不同附加滤过与数字乳腺X线摄影辐射剂量和影像质量相关性的研究[J].中华放射学杂志,2012,46:1079-1082. [5] Porter AJ,Evans EB,Foxcroft LM,et al.Mammographic and ultrasound features of invasive lobular carcinoma of the breast[J].J Med Imaging Radiat Oncol,2014,58(1):1-10. [6] Shiraishi A.Current state of digital mammography[J].Breast Cancer,2008,15:194. [7] FAO,IAEA,ILO,OECD/NEA,PAHO,WHO.IBSS.IAEA Safety series No.115[R].Vienna:1996. [8] 苗英,戎悦,章士正.乳腺摄影技术标准化的探讨[J].中华放射学杂志,2000,34(3):159-161. [9] 郭玮珍,黄伟旭,谭光享,等.乳腺X射线摄影所致受检者辐射剂量调查研究[J].中国辐射卫生,2014,23(6):502-504. [10] 郑钧正,朱志贤,唐文祥,等.乳腺X射线摄影所致受检者乳腺平均剂量调查研究[J].中华放射医学与防护杂志,2001(S1):s77-s79. [11] 张立新,王新明,杜国生.数字乳腺X线摄影中不同靶/滤过组合对影像质量的影响[J].中国医疗设备, 2014(8):126-128. [12] 刘澜涛,岳保荣.乳腺摄影中影像质量与剂量的评价研究进展[J].中华放射医学与防护杂志, 2008, 28(1):92-94. [13] 黄勇,王丹丹,何静,等.乳腺X线摄影射线剂量对乳腺X线片质量的影响[J].临床放射学杂志,2012,31(2):188-191. [14] Young K C, Burch A. Radiation doses received in the UK Breast Screening Programme in 1997 and 1998[J]. British Journal of Radiology, 2000, 73(867):278. [15] 朱维杰,王新明,冯泽臣,等.北京地区542例乳腺摄影剂量调查的分析[J].中国辐射卫生,2014,23(1):4-7. [16] 中华人民共和国国家卫生和计划生育委员会. WS 522-2017乳腺数字X射线摄影系统质量控制检测规范[S].北京:中国标准出版社,2017. |
|
|
|