Investigation of Radiation Level for Mobile Medical X-Ray Equipment by the Bedside Operation |
SHEN Qing1, CUI Qian1, JI Hao2, WANG Yu-xia3, SHEN Jia4, CHEN Biao5, ZHU Guo-yin6, CHEN Chun-hui5 |
Agency for Public Health Inspection, Xuhui Municipal Health Family Planning Commission of Shanghai, Shanghai 200237 China |
|
|
Abstract Objective In order to assist relevant authorities to build the theoretical basis for implementing targeted intervening strategies, we investigated the radiation dose rate for mobile medical X-ray equipment through bedside operation from different angles at different distance.Methods A total of 20 pieces of mobile medical x-ray equipment, from three different levels of medical institutions in District Xuhui, Shanghai, were selected in appropriate proportions and monitored.Results For mobile X-ray equipment by the bedside operation,CR is the main imaging category which accounts for 80%. A total of 75% switches adoptly remote control method,but there is 25% equipment that is operated manually. In addition, the rates did not change according to changing angles, while they declined obviously with the increase of the stationing distance. X-ray dose rate ranges in 1000~2000 μSv/h, 300~500 μSv/h,40~300 μSv/h and 10~40 μSv/h, respectively, at 1 m,2 m and 5 m distance from equipment.Conclusion For mobile medical X-ray equipment by the bedside operation, X-ray dose rate does change according to various distances, and the recommended safe distance outside should be 5 m. Appropriate lead shielding facilities like lead screen should be used in order to protect the health radiological technicians and observers.
|
Received: 14 September 2015
|
|
|
|
|
|
|
|