Measurement and Analysis of Radiation field in the Surgical Standing Areas during ERCP |
NIU Fei, YANG Shuhui, MIN Nan, ZHANG Lin, SONG Gang, ZHU Jianguo |
Institute of Radiation Medicine, Shandong Academy of Medical Science, Jinan 250062 China |
|
|
Abstract Objective To measure and evaluate the radiation dose distribution characteristics in the standing areas of the operator during endoscopic retrograde cholangiopancreatography (ERCP), and to provide radiation dose data for the protection of ERCP endoscopist.Methods The areas that ERCP operators often stay in during exposure were divided into 40 test points. Five data were measured at each test point according to different altitudes. The ambient dose equivalent rate was measured by X and gamma radiation dose rate meter.Results The ambient dose equivalent rate of the operator's standing area was 2.3~386.0 μSv/h. Generally, the closer to the center of the irradiation field, the higher the ambient dose equivalent rate around it. For the same position, the dose equivalent rate around the head, chest, abdomen and lower limb height is close, while the dose equivalent rate around the foot is slightly lower, but there were some exceptions.Conclusion The higher ambient dose rates in the standing area of ERCP surgeons are mainly obtained in the areas where the operators often stand. Therefore, effective protective measures must be taken to ensure the health and safety of the ERCP surgeons.
|
Received: 15 March 2018
|
|
|
|
|
[1] UNSCEAR.Sources and effects of ionizing radiation[R].New York:United Nations, 1993,69-71.
[2] 国家卫生和计划生育委员会.GBZ 130-2013医用X射线诊断放射防护要求[S].北京:中国标准出版社,2013.
[3] 刘长才,张琳,闵楠,等.三种常见介入诊疗中机房内的辐射场分布[J].中华放射医学与防护杂志,2014, 34(1):48-53.
[4] Campbell N,Sparrow K,Fortier M,et al.Practical radiation safety and protection for the endoscopist during ERCP[J].Gastrointestinal Endoscopy,2002,55(4):552-557.
[5] Miiler DL,Batter S,Cole PE,et al.Radiation doses in interventional radiology procedures:the RAD-lR study part l:overall measures of dose[J].J Vasc Interv Radiol, 2003,14(6):711-727.
[6] 郁鹏,尉可道,李田昌,等.介入诊疗区域内辐射场的测定与评价[J].中华放射医学与防护杂志,2005,25(3):270-271.
[7] Buls N,Pages J,Mana F,et al.Patient and staff exposure during endoscopic retrograde cholangiopancreatography[J].British Journal of Radiology,2002,75(893):435-443.
[8] Seo D,Kim KH,Kim JS,et al.Evaluation of radiation doses in patient and medical staff during endoscopic retrograde cholangiopancreatography procedures[J].Radiation Protection Dosimetry,2016,168(4):516-522.
[9] Sulieman A,Paroutoglou G,Kapsoritakis A, et al.Reduction of radiation doses to patients and staff during endoscopic retrograde cholangiopancreatography[J].Saudi Journal of Gastroenterology,2011,17(1):23-29.
[10] Sulieman A,Elzaki M,Khalil M,et al.Occupational exposure to staff during endoscopic retrograde cholangiopancreatography in Sudan[J].Radiation Protection Dosimetry,2011, 144(1/4):530-533.
[11] Larkin CJ,Workman A,Wright RE,et al.Radiation doses to patients during ERCP[J].Gut,1999,44(1):A10.
[12] Tsalafoutas IA,Paraskeva KD,Yakoumakis EN, et al. Radiation doses to patients from endoscopic retrograde cholangiopancreatography examinations and image quality considerations[J].Radiat Prot Dosimetry,2003,106(3):241-246.
[13] Naidu LS,Singhal S,Preece DE, et al. Radiation exposure to personnel performing Endoscopic retrograde Cholangiopancreatography[J].Postgrad Med J,2005,81(960):660.
[14] Jaume,Boix,Vicente,et al.Radiation dose to patients during endoscopic retrograde cholangiopan-creatography[J].World Journal of Gastrointestinal Endoscopy,2011,3(7):140-144.
[15] Larkin CJ,Workman A,Wright RE,et al.Radiation doses to patients during ERCP[J].Gastrointestinal Endoscopy, 2011,53(2):161-164.
[16] Dumonceau JM,Garcia-Fernandez FJ,Verdun FR, et al.Radiation protection in digestive endoscopy:European Society of Digestive Endoscopy(ESGE) Guideline[J].Endoscopy, 2012,44(4):408-421.
[17] Tsapaki V,Paraskeva KD,Mathou N,et al.Patient and endoscopist radiation doses during ERCP procedures[J].Radiation Protection Dosimetry,2011,147(1/2):111-113.
[18] Son BK,Lee KT,Kim JS,et al.Lack of radiation protection for endoscopists performing endoscopic retrograde cholangiopancreatography[J].The Korean journal of gastroenterology,2011,8(2):93-99.
[19] Ertel A,Nadelson J,Shroff AR,et al.Radiation dose reduction during radial cardiac catheterization evaluation of a dedicated radial angiography absorption shielding drape[J]. ISRN Cardiol, 2012,14:769167.
[20] O'Connor U,Gallagher A,MaloneL,et al.Occupational radiation dose to eyes from Endoscopic retrograde cholangiopancreatography procedures in light of the revised eye lens dose limit from the International Commission on Radiological Protection[J]. British Journal of Radiology,2013,86(1022):20120289. |
|
|
|