Investigation on quality control and radiation protection of dental X-ray equipment in Beijing Dongcheng District |
SHI Hongmei, ZHAO Sijing, WEI Xiang, LIU Jinglei |
Beijing Dongcheng district center for disease control and prevention, Beijing 100050 China |
|
|
Abstract Objective To investigate the status of quality control of dental X-ray equipment in Beijing Dongcheng District, to explore the problems involved in radiation protection.Methods 41 medical dental X-ray machines were detected in Beijing Dongcheng District in accordance with Specification for testing of quality control in dental X-ray equipment (WS 581—2017).Results The qualified rate was higher in imported machines (100%) than that in domestic machines (76.92%), the overall qualified rate was 92.68%. In performance detection, the qualified rates were 100% except for the deviation of loading time (92.68%) and tube voltage (95.12%). No leakage was detected in 41 dental institutes.Conclusion The overall status of dental X-ray machines was good in Beijing Dongcheng District. More attention was paid to shield protection than to quality control of dental X-ray machines. Therefore, health administration should take measures to strengthen advocation of WS 581—2017, and improve radiation protection through routine equipment performance test, supervision of X-ray protection and radiation health education.
|
Received: 01 April 2021
|
|
|
|
|
[1] 李军. 影响医用诊断X射线机计量性能的主要因素分析[J]. 计量与测试技术,2008,35(8):21-22. DOI: 10.3969/j.issn.1004-6941.2008.08.012 Li J. Analysis on the main impact factors of medical diagnostic X - ray measurement capability[J]. Metrol Meas Tech, 2008, 35(8): 21-22. DOI: 10.3969/j.issn.1004-6941.2008.08.012 [2] 张振庆. 全景X线牙片机辐射剂量控制的研究现状[J]. 医疗装备,2017,30(6):196-197. DOI: 10.3969/j.issn.1002-2376.2017.06.157 Zhang ZQ. Radiation dose control of dental panoramic X-ray Machine[J]. Chin J Med Device, 2017, 30(6): 196-197. DOI: 10.3969/j.issn.1002-2376.2017.06.157 [3] 中华人民共和国国家卫生和计划生育委员会. GBZ 130—2013医用X射线诊断放射防护要求[S]. 北京: 中国标准出版社, 2014. National Health Commission of the People's Republic of China.GBZ 130—2013Requirements for radiological protection in medical X-ray diagnosis[S]. Beijing: Standards Press of China, 2014. [4] 国家卫生和计划生育委员会. 牙科X射线设备质量控制检测规范: WS 581—2017[S]. 北京: 人民卫生出版社, 2017. National Health and Family Planning Commission. Specification for testing of quality control in dental X-ray equipment: WS 581—2017[S]. Beijing: People’s medical publishing house, 2017. [5] 郭火仲. 2017年福建省161台(套)医用常规X射线诊断设备质量控制检测结果与分析[J]. 中国辐射卫生,2019,28(4):404-406, 409. DOI: 10.13491/j.issn.1004-714x.2019.04.015 Guo HZ. Investigation and analysis of quality control testing of 161 sets of medical X-ray diagnostic equipment in Fujian Province in 2017[J]. Chin J Radiol Health, 2019, 28(4): 404-406, 409. DOI: 10.13491/j.issn.1004-714x.2019.04.015 [6] 李笑鑫, 杨凯川, 李松, 等. 部分牙科X射线机防护性能的检测结果与分析[J]. 中国辐射卫生,2015,24(6):636-638. DOI: 10.13491/j.cnki.issn.1004-714x.2015.06.026 Li XX, Yang KC, Li S, et al. Results and analysis of protective performance of some dental X-ray machines[J]. Chin J Radiol Health, 2015, 24(6): 636-638. DOI: 10.13491/j.cnki.issn.1004-714x.2015.06.026 [7] 郑海英, 温小庭, 罗小铭, 等. 珠三角某市牙片机防护性能调查分析[J]. 中国辐射卫生,2016,25(2):208-209. DOI: 10.13491/j.cnki.issn.1004-714x.2016.02.029 Zheng HY, Wen XT, Luo XM, et al. Analysis of protection performance of dental X-ray machines in Chu Chiang Delta[J]. Chin J Radiol Health, 2016, 25(2): 208-209. DOI: 10.13491/j.cnki.issn.1004-714x.2016.02.029 [8] 白鹏. 分析口腔X射线机影像质控检测[J]. 影像技术,2015,27(4):53-54, 60. DOI: 10.3969/j.issn.1001-0270.2015.04.25 Bai P. Analysis of image quality control of dental X-ray machines[J]. Image Technol, 2015, 27(4): 53-54, 60. DOI: 10.3969/j.issn.1001-0270.2015.04.25 [9] 杨伟华. 北京市海淀区牙科X射线机调查分析[J]. 中国辐射卫生,2006,15(3):333-335. DOI: 10.13491/j.cnki.issn.1004-714X.2006.03.053 Yang WH. Research of dental X-ray machines in Beijing Haidian district[J]. Chin J Radiol Health, 2006, 15(3): 333-335. DOI: 10.13491/j.cnki.issn.1004-714X.2006.03.053 [10] 安欣华, 张德钦, 刘华京. 北京市医用口腔X射线机影像质控检测与分析[J]. 首都公共卫生,2014,8(5):205-207. DOI: 10.16760/j.cnki.sdggws.2014.05.009 An XH, Zhang DQ, Liu HJ. Quality control testing and analysis on image of medical dental X-ray machine in Beijing[J]. Cap J Public Heal, 2014, 8(5): 205-207. DOI: 10.16760/j.cnki.sdggws.2014.05.009 [11] 魏琴, 谢贤宇, 郑森兴, 等. 医院新型放射防护管理实践与体会[J]. 中国辐射卫生,2019,28(2):179-181, 185. DOI: 10.13491/j.issn.1004-714x.2019.02.018 Wei Q, Xie XY, Zheng SX, et al. Practice and experience of a new radiological protection management model in hospital[J]. Chin J Radiol Heal, 2019, 28(2): 179-181, 185. DOI: 10.13491/j.issn.1004-714x.2019.02.018
|
|
|
|