Research on temperature peak-shift correction methods for NaI (Tl) gamma spectrum |
FANG Dengfu, WEI Yingjing, CUI Wei, FENG Mei, TANG Zhihui, LI Zhigang, YI Hengguan |
China Institute for Radiation Protection, Taiyuan 030006 China |
|
|
Abstract Objective The experiment project was designed to explore the variation of NaI (Tl) gamma spectrometer channels with environmental temperature. 60Co and 152Eu were used to verify the reliability of the correction methods.Methods Two correction methods were applicated, which were curve fitting correction method and known measurement peak correction method.Results The experimental results showed that temperature changes had an effect on NaI (Tl) measured spectra peak. The relative peak will shift to the right at 5℃ by 9.6%, and to the left at 60℃ by 16%, with the reference temperature set at 25℃. The two methods are based on the channel change due to temperature changes, and they could effectively correct the temperature peak-drift.Conclusion In order to make the measured spectrum information accurate and reliable in field monitoring, it is suggested to monitor the environmental temperature so as to correct the measured data.
|
Received: 10 July 2020
|
|
|
|
|
[1] Milbrath BD, Choate BJ, Fast JE, et al. Comparison of LaBr3: Ce and NAI(Tl) scintillators for radio-isotope identification devices[J]. Nucl Instruments Methods Phys Res Sect A: Accel Spectrometers Detect Assoc Equip, 2007, 572(2): 774-784. DOI: 10.1016/j.nima.2006.12.003 [2] Moszyński M, Nassalski A, Syntfeld-Każuch A, et al. Temperature dependences of LaBr3(Ce), LaCl3(Ce) and NaI(Tl) scintillators[J]. Nucl Instruments Methods Phys Res Sect A: Accel Spectrometers Detect Assoc Equip, 2006, 568(2): 739-751. DOI: 10.1016/j.nima.2006.06.039 [3] 姚杰, 钱爱君, 高林峰. 2018年参加全国放射性核素γ能谱分析比对结果[J]. 中国辐射卫生,2019,28(3):321-324. DOI: 10.13491/j.issn.1004-714x.2019.03.028 Yao J, Qian AJ, Gao LF. Results of participating in the national intercomparison of radionuclideanalysis by γ spectroscopy in 2018[J]. Chin J Radiol Heal, 2019, 28(3): 321-324. DOI: 10.13491/j.issn.1004-714x.2019.03.028 [4] 屠琦, 陈乐庚. 基于DSP的NaI(Tl)闪烁探测器温漂补偿方法[J]. 测控技术,2016,35(7):13-16. DOI: 10.19708/j.ckjs.2016.07.004 Tu Q, Chen LG. A compensation method of temperature drift of NaI(TI) scintillation detector based on DSP[J]. Meas Control Technol, 2016, 35(7): 13-16. DOI: 10.19708/j.ckjs.2016.07.004 [5] 李文杰, 侯霞, 郝秋伟. 碘化钠γ谱仪温度效应对陶瓷材料放射性核素测定的影响研究[J]. 中国测试,2015,41(4):19-23, 28. DOI: 10.11857/j.issn.1674-5124.2015.04.005 Li WJ, Hou X, Hao QW. The study of temperature effect of sodium iodide γ-ray detector on measurement of ceramic materials radionuclide[J]. Chin Meas Test, 2015, 41(4): 19-23, 28. DOI: 10.11857/j.issn.1674-5124.2015.04.005 [6] 杨焕章. 航空γ能谱探头温度效应及校正技术研究[D]. 成都: 成都理工大学, 2009. Yang HZ. Temperature character analysis and adjustment technique for the detector of aerial gamma-ray spectrometric[D]. Chengdu: Chengdu University of Technology, 2009. [7] Williams D, Snelling GF, Pickup J. A γ-spectrum stabiliser with compensation for the effects of detector temperature variation[J]. Nucl Instruments Methods, 1966, 39(1): 141-149. DOI: 10.1016/0029-554X(66)90054-1 [8] Casanovas R, Morant JJ, Salvadó M. Temperature peak-shift correction methods for NaI(Tl) and LaBr3(Ce) gamma-ray spectrum stabilisation[J]. Radiat Meas, 2012, 47(8): 588-595. DOI: 10.1016/j.radmeas.2012.06.001 [9] 方登富, 李强, 唐智辉, 等. 海水就地γ谱仪能谱温漂修正研究[J]. 核电子学与探测技术,2017,37(6):564-568. DOI: 10.3969/j.issn.0258-0934.2017.06.002 Fang DF, Li Q, Tang ZH, et al. Study on temperature drift correction of energy spectrum from seawater in-situ γ-ray spectrometer[J]. Nucl Electron Detect Technol, 2017, 37(6): 564-568. DOI: 10.3969/j.issn.0258-0934.2017.06.002 |
|
|
|