Experimental research on partial dosimetric performance for eye lens dosimeter and ring dosimeter |
DING Yanqiu, GUO Wen, HU Aiying, WANG Kaiyi |
Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100086 China |
|
|
Abstract Objective To compare the test results of domestic and imported eye lens and ring dosimeters, and master some dosimetric performance indexes of eye lens and ring dosimeters.Methods According to IEC 62387—2012 and GBZ 128—2019, the performance of the eye lens dosimeters and ring dosimeters with the domestic thermoluminescence detectors were compared with the imported thermoluminescence detectors in photon response. The dose response deviation of the imported eye lens dosimeter and imported ring dosimeter was studied under the same irradiation condition. The difference of linear response between imported ring dosimeter and domestic ring dosimeter was studied under the same irradiation condition.Results The same dose was irradiated to domestic and imported thermoluminescence detectors in the imported eye lens and imported ring dosimeters, and the linear correlation coefficient all met R2 > 0.999. The maximum response deviation of the imported eye lens dosimeter and ring dosimeter was less than 10.0% for the same dose. Two kinds of imported ring dosimeters and one kind of domestic ring dosimeter were irradiated with the same dose, and the linear correlation coefficient all met R2 > 0.99.Conclusion Both domestic detectors and imported detectors can meet the needs of monitoring requirements. There was no obvious difference in linear response between imported ring dosimeter and domestic ring dosimeter. The repeatability of imported eye lens dosimeter and imported ring dosimeter met the performance requirements of the dosimeter.
|
Received: 14 June 2020
|
|
|
|
|
[1] 宋颖, 欧向明, 刘辉, 等. 核医学职业人员眼晶状体剂量测量方法研究[J]. 中国医学装备,2019,16(10):25-27. DOI: 10.3969/J.ISSN.1672-8270.2019.10.007 Song Y, Ou XM, Liu H, et al. Study on the measurement method for the dose of eye lens of nuclear medical staff[J]. Chin Med Equip, 2019, 16(10): 25-27. DOI: 10.3969/J.ISSN.1672-8270.2019.10.007 [2] International Commission on Radiological Protection (ICRP) Statement on Tissue Reactions[R]. ICRP, 2011. [3] 中华人民共和国国家卫生健康委员会. GBZ 128—2019职业性外照射个人监测规范[S]. 北京: 中国标准出版社, 2019. National Health Commission of the People's Republic of China. GBZ 128—2019 Specifications for individual monitoring of occupational external exposure[S]. Beijing: Standard press of China, 2019. [4] 国家核行业标准. EJ/T 1178—2015指端和眼睛肢端和眼睛热释光个人剂量计[S]. 北京:国防科学技术委员会, 2005. National Nuclear Industry Standard.EJ/T 1178—2015 Individual thermoluminescencedosemeters for extremities and eyes [S].Beijing: Commission of Science and Technology for National Defence, 2005. [5] International Organization for Standardization.X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy-Part 3: Calibration of area and personal dosemeters and the measurement of their response as a function of energy and angle of incidence[R]. ISO, 2019. [6] IEC.Radiation protection instrumentation -Passive integrating dosimetry systems for personal and environmental monitoring of photon and beta radiation[R].IEC,2012. [7] 李海亮, 王仲文, 许家昂, 等. 眼晶状体Hp(3)剂量计的医用诊断X射线辐射质校准研究[J]. 中国辐射卫生,2016,25(4):385-388. DOI: CNKI:SUN:REDI.0.2016-04-002 Li HL, Wang ZW, Xu JA, et al. Study of Calibration the dosemeter for measuring eye lens dose in terms of Hp(3) with medical diagnostic X ray[J]. Chin J Radiol Heal, 2016, 25(4): 385-388. DOI: CNKI:SUN:REDI.0.2016-04-002 |
[1] |
. [J]. , 2015, 24(4): 404-407. |
|
|
|
|