The influence of the detection window wrapped with plastic film on the surface contamination detection |
CHEN Zhongmin, HU Hongfang, LI Wei, DONG Suhe, YU Huijie, WANG Sinian, LV Xiaoli, DENG Zhirong |
Rocket Force Characteristic Medical Center, Beijing 100088 China |
|
|
Abstract Objective To study the influence of the detection window of the surface contamination detector wrapped with plastic film on the surface contamination detection.Methods RDA150 and Como170 type surface contamination detectors were selected to measure the surface α-ray contamination level of Americium-241 source under the condition of no film, single-layer film and double-layer film covering the observation window of the equipment, respectively.Results Under the condition of no film, the average measurement result of RDa150 surface contamination detector's 5 measurements was 228.01 Bq/cm2, and that of Como170 type surface contamination detector was 231.05 Bq/cm2; Under the condition of single-layer film covering, the average measurement results of RDa150 type surface contamination detector's 5 measurements was 96.34 Bq/cm2, and the Como170 surface contamination detector were 100.31 Bq/cm2 which decreased by about half. Under the condition of double-film coverage, the average measurement results of RDa150 type surface contamination detector's 5 measurements were 0.45 Bq/cm2, and the Como170 type surface contamination detector were 0.65 Bq/cm2.Conclusion When the surface contamination level is high, a single layer of plastic film can be used to cover the equipment detection window to reduce the risk of equipment contamination. When the decontamination is completed or the level of surface contamination is low, the equipment detection window cannot be wrapped with plastic film.
|
Received: 09 May 2020
|
|
|
|
|
[1] 刘英. 依法履行职责, 做好核应急和放射应急医学救援[J]. 中华放射医学与防护杂志,2008(4):426-428 [2] 唐木涛, 高伟. 军队辐射防护与管理[M]. 北京: 军事医学科学出版社, 2014: 204-206. [3] 江海洋. 核应急准备中有关问题的探讨与思考[J]. 中国辐射卫生,2005,14(2):107-108 [4] 许荩, 钟军, 郭翔博, 等. 影响表面污染仪测量准确度的因素研究[J]. 科技视界,2016,6(13):29-30 [5] 中央军委后勤保障部卫生局. 军队核应急医学救援流程与操作规范(2017版)[M].2017. [6] International Atomic Energy Agency. Radiation Monitoring Equipment[S].2009. [7] 戴定, 伍浩松. 全球核电装机容量2019年小幅降低[J]. 国外核新闻,2020(1):23-25 [8] 李锦, 唐丽丽, 喻正伟, 等. 福岛核事故后核电厂应急监测技术改进[J]. 核电子学与探测技术,2015,35(10):1038-1042 [9] 刘书锋, 刘承军, 马得勋, 等. 便携式单兵核应急防护背囊的研制[J]. 中国辐射卫生,2018,27(4):357-359 [10] 高峰. 核辐射卫生防护[M]. 北京: 人民军医出版社, 2011. [11] 李文红, 拓飞, 周强, 等. 核辐射应急情况下放射性核素的γ能谱快速分析方法[J]. 中国辐射卫生,2019,28(6):684-687
|
[1] |
YANG Shuhui, LIU Qian, SONG Gang, LI Hailiang, NIU Fei, CHEN Rui, MIN Nan, LIU Fang, CHEN Yingmin. Investigation and analysis of radiotherapy and nuclear medicine exposure frequency of 13 hospitals in Shandong Province in 2016[J]. , 2020, 29(6): 586-589. |
|
|
|
|