Measurement and comparison of intraocular lens dose in five cranial CT scanning methods |
YANG Hongyun1,2,3, LI Hailiang2, YANG Shuhui2, ZHANG Lin2, MIN Nan2, ZHU Jianguo2 |
1. School of Medicine and Life Sciences, University of Jinan-Shandong Academy of Medical Sciences, Jinan 250200 China; 2. Institute of Radiation Medicine, Shandong Academy of Medical Sciences; 3. Xintai Affiliated Hospital of Taishan Medical University(Xintai People's Hospital) |
|
|
Abstract Objective Measure and compare the dose of ocular lens in five cranial CT scanning methods by using simulation human body model and thermoluminescence dosimeter (TLD).Methods The TLD was placed in the reserved hole of the eye lens of the head model.Five cranial CT scanning methods were performed on 8 CT. At the end of each scan, take down the TLD was taken down,andrecord the numbered for later measurement.Results The eye lens radiation dose measured of anthropomorphic phantom by TLD under the condition of head CT scanning respectively was that (17.78±10.03)mGy under the brainhead CT scanning,(17.45±7.92)mGy under the Orbital cavity CT scanning,(14.65±3.80)mGy under the nasal bone CT scanning,(9.93±6.25)mGy under the Sinuses CT scanning,(7.94±5.18)mGy under the temporal bone CT scanning.Conclusion The overall level of the exposure dose to the intraocular lens during the CT scan of the skull was(13.55±7.89)mGy. The dose of ocular lens shows significant difference in different CT and craniocerebral scanning methods.
|
Received: 10 January 2019
|
|
|
|
|
[1] 伍健,耿建华.CT设备质量控制的现状与展望[J].中国医学装备,2008,15(11):163-166. [2] 郑钧正.医学影像学的时代重任[J].医学研究杂志,2015, 44(5):1-4. [3] 苏垠平.部分地区X射线诊断照射频度调查及CT所致癌症风险的研究[D].北京:中国疾病预防控制中心,2014. [4] 张代斌.重庆市放射工作人员317名眼晶状体状况分析[J].中国当代医药,2010,17(2):122-123. [5] ICRP.ICRP Publication102 Managing Patient Dose in Mult-detector Computed Tomography[M]. Oxford:Pergamon Press,2007. [6] 白光,韩华锋.我国职业外照射个人剂量检测的回顾与思考[J].辐射防护通讯,2009,29(4):1-7. [7] 中华人民共和国卫生部.WS/T 391-2012 CT检查操作规程[S].北京:中国标准出版社,2012. [8] 李士骏.电离辐射剂量学基础[M].苏州:苏州大学出版社,2008. [9] GBZ 207-2016, 外照射个人剂量系统性能检验规范[S].北京:中国标准出版社,2016. [10] 韩志伟,赵风玲,陈玉浩,等.职业性放射性白内障的临床报告及诊断探讨[J].中华放射医学与防护杂志,2013,33(4):421-422. [11] Worgul B V, Kundiyev Y I, Sergiyenko N M, et al. Cataracts among Chernobyl clean-up workers:implications regarding permissible eye exposures[J]. Radiation Research, 2007, 167(2):233-243. [12] International Commission on RadiologicalProtection(ICRP).Statement on tissue reactiongs[R].ICRP, 2011. [13] Yeoman J, Howarth L, Britten A, et al. Gantry angulation in brain CT:dosage implications, effect on posterior fossa artifacts, and current international practice[J]. Radiology, 1992, 184(1):113-116. [14] 殷志杰,张显鹏,李祥林,等.热释光探测器在头部CT检查中监测晶状体器官的辐射剂量[J].中国介入影像与治疗学,2014,11(5):321-323. |
|
|
|