Study on the calibration method of absorbed dose to water using pinpoint ionization chamber |
HE Zhijian, YUAN Jilong, XUE Xian, CHENG Jinsheng |
Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088 China |
|
|
Abstract Objective To establish a calibration method of the absorbed dose to water using pinpoint ionization chamber at the radiation quality of 60CO gamma rays. Methods Reference dosimeter (FC65-G Farmer ionization chamber with DOSE 1 electrometer) was calibrated by the China Institute of Metrology for obtaining the absorbed dose calibration factor to water. 60Co gamma rays and the IAEA TRS-398 measuring procedure were used to measure the absorbed dose at depth of 10 cm water with the reference dosimeter. Alternative method, the water absorbed dose measurement and calibration of two pinpoint ionization chamber dosimeters were carried out. Replacing the 60Co gamma radiation field, the reference dosimeter and two pinpoint ionization chamber dosimeters were used to verify the dose. Results The measured results of absorbed dose to water at depth of 10 cm water at reference conditions were 0.249 9 Gy for reference dosimeter, 0.248 0 Gy and 0.250 0 Gy for two pinpoint ionization chambers,respectively. The relative deviations of the measurement were within ±0.8% between the reference dosimeter and two pinpoint ionization chamber dosimeters. The measurement uncertainty of the pinpoint ionization chamber was 2.8% (k=2). Conclusion The pinpoint ionization chamber can be used to measure the absorbed dose to water of small field.
|
Received: 04 September 2018
|
|
|
|
|
[1] 谢耩,王运来,马林,等.电离室灵敏体积对调强放射治疗剂量学验证准确性的影响[J].中华放射医学与防护杂志,2007,27(6):582-583. [2] 蔡汉飞,裴元吉,沈学明,等.高能光子相对输出因子的测量与分析[J].中国医学物理学杂志,2009,26(4):1273-1276,1290. [3] 胡传朋,程晓军,乔红兵,等.TLD核查6 MV光子线束剂量学参数的方法验证[J].中国辐射卫生,2014,23(1):23-25. [4] 李勤,李光俊,韩军,等.加速器小射野剂量学特性的临床研究[J].中国辐射卫生,2007,16(4):444-445. [5] 李光俊,李勤,韩军,等.调强放疗中小野剂量学特性研究[J].中华放射肿瘤学杂志,2007,16(6):469-472. [6] 何志坚,罗素明,高磊,等.头部X刀焦点输出剂量测量方法研究[J].中华放射医学与防护杂志,2009,29(3):315-316. [7] 姚杏红,陈立新,靳光华.不同指形电离室测量高能光子束吸收剂量的比较[J].中华放射医学与防护杂志,2012,32,(4):412-415. [8] 张晓军,楚建军,张福正,等. 利用0.13cc电离室对头颈部肿瘤调强放疗计划的剂量学验证[J].现代肿瘤医学,2012.20(7):1445-1447. [9] 赵进沛,张富利,王雅棣,等.TomoTherapy治疗计划剂量验证方法的研究[J].医疗卫生装备,2013,34(7):105-107. [10] 罗琛,范耀东,王焕宇,等.三种电离室在小野条件下的剂量测量性能研究[J].计量技术,2018(2):9-12. [11] International Atomic Energy Agency. IAEA TRS 398.Absorbed dose Determination in External Beam Radiotherapy:An International Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water[R].Vienna:IAEA, 2000. [12] International Atomic Energy Agency. IAEA TRS No. 469. Calibration of Reference Dosimeters for External Beam Radiotherapy[R]. Vienna:IAEA, 2009. [13] 程金生,路吉龙,刘雅,等.三种放疗剂量学规程在60Co辐射场中测量对比研究[J].中国辐射卫生,2016,25(5):513-515,519. |
|
|
|