The feasibility study of ATCM combined with ASiR-V technique to detect the pulmonary GGO nodules with a diameter around 10 mm using ultra-low dose CT |
Wang Naiwu1, Han Wenjuan2, Li Qingguo1, Li Ying1, Yang Chenxiao1, Jia Shouqiang1, Wang Junxin1 |
1. department of radiology, Laiwu Affiliated Hospital of Taishan Medical University, Laiwu 271100 China; 2. Scientific education section, Laiwu Affiliated Hospital of Taishan Medical University |
|
|
Abstract Objective To investigate the feasibility of ATCM combined with ASiR-V technique for ultra-low-dose CT examination and to determine the optimal level of ASiR-V for displaying pulmonary ground-glass-opacity nodules (GGO) with a diameter around 10mm.Methods Three GGO of diameter 12 mm, 10 mm and 8mm were randomly placed in a chest phantom. The chest phantom was scanned on a Revolution CT and ATCM at a high noise index(NI=35 Hu), with different levels of ASiR-V (0%、20%、40%、60%、80%、100%). Scanning parameters included:tube voltage 120 kVp,Helical scan,GP:0.5s,SPR:0.992:1. Images were reconstructed by standard algorithm with slice thickness of 5mm. The background image noise of the three nodules was measured as the average of the standard deviations of ROIs (≧1.0 cm2) placed in the heart as the same slice. CT dose index (CTDIvol) and dose length product (DLP) were measured and effective radiation dose (ED) was calculated. Objective analysis and subjective evaluation was done with image quality and radiation dose. Image noise with different nodules at different weighted levels was compared with Single factor ANOVA and correlation analysis by using SPSS 20.0 statistical software. Subjective evaluation was performed by two radiologists using a 5-point scale, and Kappa test was used to examine the agreement of the two readers.Results NI=35Hu.With the increase of prescribed ASiR-V weight, the tube current decreased(16~34 mA,11~24 mA,9~15 mA,9~10 mA,9 mA,9 mA),and ED decreased (0.62 mSv,0.44 mSv,0.30 mSv,0.25 mSv,0.24 mSv,0.24 mSv). The image background noises (SD) at the slice of the three nodules (12mm,10mm,8mm) were (0% ASiR-V,21.33±1.88;20%ASiR-V,21.27±1.43;40% ASiR-V,19.30±1.90;60%ASiR-V,13.73±1.36;80%ASiR-V,10.63±0.45;100%ASiR-V,9.70±0.82),the SD decreased(P<0.05). Two physicians had good agreement on subjective scores (K=0.75). The subjective scores with the increase of ASiR-V were 4.50、4.50、4.67、5.00、4.16、3.50. The highest subjective score was with 60% ASiR-V.Conclusion Radiation dose and image noise can be significantly reduced by using ATCM combined with ASiR-V technique on Revolution CT, and Image quality meets the requirement of clinic diagnosis. Ultra-low-dose scanning using NI=35 combined with 60% ASiR-V is the optimal scanning parameters for the GGO nodules with a diameter around 10 mm.
|
Received: 13 July 2018
|
|
|
|
|
[1] 白春学.应重视和加强对肺癌早期诊断及预后的研究[J].中华医学杂志,2013,93(38):3009-3010. [2] Brenner DJ,Hall EJ.Computed tomography-an increasing source of radiation exposure[J].N Engl J Med,2007,357(22):2277-2284. [3] Chang B,Hwang JH,Choi YH,et al.Natural history of pure ground-glass opacity lung nodules detected by low-dose CT scan[J].Chest,2013,143(1):172-178. [4] 周莹,刘士远.美国国立综合癌症网肺癌筛查指南解读(2012)[J].肿瘤影像学,2013,22(4):331-335. [5] Naidich DP,Bankier AA,MaeMahon H,et al.Recommendations for the management of subsolid pulmonary nodules detected at CT:A statement from the Fleischner Society[J].Radiology,2013,266(1):304-317. [6] Zhang XH,Yang L,wu J,et al.Diagnostic accuracy and its affecting factors of dual-source CT for assessment of coronary stents pateney and in-stent restenosis[J].Chin Mad J(En91),2012,125(11):1936-1940. [7] Jang nJ,Lee KS,Kwon OJ,et al.Bronchioloalveolar carcinoma:focal area of ground-glass attenuation at thin-section CT as an early sign[J].Radiology,1996,199(2):485-488. [8] Suzuki K,Kusumoto M,Watanabe S, et al.Radiologic classification of small adenocarcinoma of the lung:radiologic-pathologic correlation and its prognostic impact[J].Ann Thorac Surg,2006,81(2):413-419. [9] Nognchi M, Morikama A,Kawasaki M, et al. Small adenocarcinoma of the lung. Histologic characteristics and prognosis[J].Cancer,1995,75(12):2844-2852. [10] Brenner DJ.Radiation risks potentially associated with low-dose CT screening of adult smokers for lung cancer[J].Radiology, 2004,23l(2):440-445. [11] Kodama K,Higashiyama M,Yokouchi H,et al.Natural history of pure ground-glass opacity after long-term follow-up of more than 2 years[J].Ann Thorac Surg,2002,73(2):386-392. [12] Takashima S,Maruyama Y,Hasegawa M,et al.CT findings and progression of small peripheral lung neoplasms having a replacement growth pattern[J].AJR Am J Roentgenol,2003, 180(3):817-826. [13] Kaneda H,Nakano T,Taniguchi Y,et al.A decrease in the size of ground glass nodules may indicate the optimal timing for curative surgery[J].Lung Cancer,2014,85(2):213-217. [14] Kubo T,Lin P J,Stiller W,et al.Radiation dose reduction in chest CT:a review[J].AJR,2008,190(2):335-343. [15] Ko J P,Rusinek H,Naidich D P,et al.Wavelet compression of low-dose chest CT data:effect on lung nodule detection[J].Radiology,2003,228(1):70-75. [16] 徐列印.CT低剂量技术研究进展[J].华夏医学,2017,30(1):172-175. [17] HUNDT W,RUST F P,STABLER A M,et al.Dose reduction in multilice computed tomography[J].J Computed Assist Tomogram,2005,29(1):140-147. [18] KALRA MK,MACHER MM,KAMATH RS,et al.Sixteen-detector row CT of abdomen and pelvis:study for optimization of Z-axis modulation technique performed in 153 patients[J].Radiology,2004,233(1):241-249. [19] BRISSE HJ,MADEC L,GABORIAUD G, et al. AutomAtic exposure control in multichannel CT with tube current modulation to achieve a constant level of image noise:experimental assessment on pediatric phantoms[J].Med Physics,2007,34(7):3018-3033. [20] Liu YJ,Zhu PP,Chen B,et al.A new iterative algorithm to reconstruct the refractive index[J].Phys Med BioI,2007,52(12):L5-L13. [21] Kilic K,Erbas G,Guryildirim M, et al.Lowring the dose in head CT using adaptive iterative reconstruction[J].AJNR Am J Neuroradiol,2011,32(9):1578-1582. [22] 刘琴,侯阳,郭启勇.迭代重建技术在低剂量CT中的应用进展[J].中国临床医学影像杂志,2013,24(11):805-807. [23] 陆锐,何健,周科峰,等.自动管电流调制扫描及ASIR重建下胸部CT噪声指数的优化[J].医学影像学杂志,2016,26(6):1012-1016. [24] 肖慧娟,侯平,刘杰,高剑波,等.低剂量CT胸部扫描最佳噪声指数及适应迭代重建权重的选择分析[J].中华医学杂志,2015,95(37):3041-3044. |
|
|
|