Estimation of Dual-Source CT Effective Dose |
WANG Qiang, FU Qiang, LIN Lin, ZHU Jian-quan |
Changzhou CDC, Changzhou, 213022 China |
|
|
Abstract Objective To estimate dual-source CT effective doses under four different examination protocols.Methods After putting TLDs in the Alderson Radiation Therapy Phantom,we scanned the phantom under two different examination protocols.Then effective doses were calculated with the acquired absorbed doses of organs and tissues. Tissue weighting factors recommended by the ICRP were adopted in the calculation. Effective doses were also calculated using the AAPM recommended method by which the DLP was multiplied by the K factor.Results Effective doses under two different examination protocols were obtained.The effective doses were 3.66 and 2.98mSv respectively when phantom were used. The effective doses were 2.81 and 1.55 mSv respectively calculated using the K value recommended by AAPM and were all lower than that obtained using phantom.Conclusion Under the head scan mode,the absorbed dose of crystalline lens is high.Effective doses calculated using the method recommended by AAPM were lower than those using the Alderson Radiation Therapy Phantom.
|
Received: 23 August 2016
|
|
|
|
|
[1] ICRP.Managing patient dose in multi-detector computed tomography(MDCT)[R].ICRP Publication 102.Oxford:Pergamon Press,2007:34. [2] 李敏,郑理华,张晨,等.双源CT图像质量与辐射剂量的关系与分析[J].医疗卫生装备,2013,34(12):88-89. [3] 李剑,宦怡,印弘,等.低浓度对比剂和超低辐射剂量在双源CT冠状动脉成像中的应用[J].实用放射学杂志,2015,31(2):300-304. [4] ICRP.The 2007 Recommendations of the International Commission on Radiological Protection[R].ICRP Publication 103.2007:63-70. [5] DeMarco JJ,Cagnon CH,Cody DD,et a1.Estimating radiation doses from mulltidetector CT using Monte Carlo simulations:effects of different size voxelized patient models on magnitudes of organ and effective dose[J].Phys Med Biol,2007,52(9):2583-2597. [6] 张琳,朱建国,闵楠,等.3种常见介入诊疗中放射工作人员有效剂量的估算[J].中华放射医学与防护杂志,2011,,3(4):393. [7] 李士俊.电离辐射剂量学基础[M].苏州:苏州大学出版社,2008:43. [8] McCollough C,Cody D,et al.The Measurement,Reporting,and Management of Radiation Dose in CT[R]:Report of AAPM Task Group 23 of the Diagnostic Imaging Council CT Committee.USA:AAPM.2008:13. [9] 李士俊.电离辐射剂量学基础[M].苏州:苏州大学出版社,2008:31. [10] Alkadhi H,Stolzmann P,Desbiolle L,et al.Low-dose,128-slice,dual-source CT coronary angiography:accuracy andradiation dose of the high-pitch and the step-and-shoot mode[J].Heart.2010,96(12):933-938. [11] Han BK,Grant KL,Garberich R,et al.Assesment of an iterative reconstruction algorithm(SAFIRE) on image quality in pediatric cardiac CT datasets[J].J Cardiovasc Comput Tomogr,2012,6(3):200-204. [12] DeMarco JJ,Cagnon CH,Cody DD,et a1.Estimating radiation doses from mulltidetector CT using Monte Carlo simulations:effects of different size voxelized patient models on magnitudes of organ and effective dose[J].Phys Med Biol,2007,52(9):2583-2597. [13] Wall BF.Radiation protection dosimetry for diagnostic radiology patients[J].Radiat Prot Dosim,2004,109(4):409-419. [14] Zelikman M, Kruchinin S, Burashov V. How to Estimate Multislice CT Effective Dose More Correctly[J]. Health Phys,2013,105(6):534-539. |
|
|
|