Assessment of the Eye Lens dose to Working Operator in Five Types of Interventional Fluoroscopic Procedures |
XUE xian, WEI Ke-dao, ZHAO Hong-feng, XU Hui, WANG Jian-chao, YUE Bao-rong |
National Institute for Radiological Protection, China CDC, Beijing 100088 China |
|
|
Abstract Objective To report estimated radiation dose to the eye lens of the interventionalist from five procedures.Methods The radiation doses to the eye lens for five interventional radiology fluoroscopic systems were measured by using OSLs(Optically Stimulated Luminescence) dosimetry calibrated in Hp(0.07).For the eyes, the OSLs are positioned to the right,left and middle sids. Meanwhile,fluoroscopy times,accumulated doses,and DAPs(Dose-Area Product) et.al were recorded for statistical analysis.Results The annual eye lens doses estimated for the operator with the highest workload according to the measurements and the system's log book were 15.12 mSv in CA(Coronary Angiography), 11.59 mSv in PTCA (Percutaneous Transluminal Coronary Angioplasty) and PCI(Percutaneous Coronary Intervention),8.64 mSv in Cerebral aneurysm,36.10 mSv in Pacemaker Therapy,12.78 mSv in Radiofrequency ablation. In Pacemaker Therapy the dose of equivalent of eye lens was the highest in he five procedures.The eye lens doses were different in different procedures,fluoroscopy times,accumulated doses and DAPs.Conclusion The eye lens dose of operators in Pacemaker Therapy will receive an annual equivalence dose of more than 20 mSv as recommended by the ICRP provided by the current study.Attention should be paid to the protection of the eyes of operators.
|
Received: 22 May 2017
|
|
|
|
|
[1] 马明强,孙培芝,孙扣红.介入放射学操作人员受照剂量及其防护监测评价[J].职业与健康,2008,19(8):14-16. [2] ICRP.Statement on tissue reaction[J/OL].ICRP ref 4825-3093-1464. http://www.icrp.org/page.asp?id=123.ICRP Statement on Tissue Reactions.pdf. [3] 刘长安,李小娟,高玲.辐射诱发白内障阈剂量判断:变化在即[J].中华放射医学与防护杂志,2011,31(3):249-251. [4] Carinous Eleftheria,Domienik Joanna, Donadille Laurent et.al. Report on the analysis of the measurement and simulation results, with an evaluation of the doses and radiation protection measures[J/OL]. http://www.oramed-fp7.eu/e n/Deli verables.D13v5.pdf. 2010-12. [5] 彭建亮,娄云,冯泽臣.3种介入术中工作人员的辐射剂量水平分析[J].中华放射医学与防护杂志,2011,31(4):395-397. [6] 潘自强,周永增,周平坤等译校.国际放射委员会2007年建议书[M].北京:原子能出版社,2008:330-332. [7] Gerber T.Dosimetry for the Lens of the Eye,Applications for Medical Staff Involved in Interventional Radiology Procedures.Lund[M].Lund University,Sweeden,2010. [8] Behrens R and Dietze G. Monitoring the eye lens:which dose quantity is adequate[J]. Phys Med Biol, 2010,55(14):4047-4062. [9] Koukoraca C,Carinou E,Simantirakis G,et al.Dose to perators during interventional radiology procedures:focus on eye lens and extremity dosimetry[J].Radiat Prot Dosim,2011,144(1-4):482-486. [10] Gualdrini G,Bordy JM,Daures J,et al.Air kerma to Hp(3) convention coefficients for photons from 10keV to 10MeV,calculated in a cylindrical phantom[J].Radiation Protection Dosimetry,2012:1-5. [11] Vanhavere F,Carinou E,Domienik J,et al.Measurments of eye lens doses in interventional radiology and cardiology:final results of the oramed project[J].Radiation Measurment,2011,46(11):1243-1247. [12] Raymond H,Lawrence T,Joaquin P,et al.Comparing Strategies for Operator Eye Protection in the Interventional Radiology Suit[J].J Vasc Interv Radiol 2010,21(11):1703-1707. [13] Kwang Pyo Kim,Donald L.Miller.Minimising radiation exposure to physicians performing fluoroscopically guided cardiac catheterization procedures:a review[J].Radiation Protection Dosimetry,2009,133(4):227-233. [14] Donald L Miller,Eliseo Vañó, Gabriel Bartal, et al, Occupational Radiation Protection in Interventional Radiology:A Joint Guideline of the Cardiovascular and Interventional Radiology Society of Europe and the Society of Interventional Radiology[J]. Cardiovasc Intervent Radiol.2010, 33(2):230-239. [15] Sturchio GM,Newcomb RD,Molella R,et al.Protective eyewear selection for interventional fluoroscopy[J].Health Phys,2013,104(Supplement 1):S11-S16. |
|
|
|