Study on the effective dose of CT in PET/CT on automatic tube current modulation |
WU Yitian1,2, GENG Jianhua1,2, DU Zhaomeng1, BI Gaochang1, QI Yonghe1, ZHANG Chaokun1, ZHENG Rong1,2, WU Ning1,2 |
1. National Cancer Center/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Department of PET-CT Center, Beijing 100021 China; 2. Department of Nuclear Medicine |
|
|
Abstract Objective To investigate the relationship between effective dose and tube current threshold, as well as noise index of CT in PET/CT, and to provide theoretical basis for optimal acquisitions through the mode of automatic tube current modulation (ATCM).Methods Two sets of PET/CT (Discovery ST-16 and Discovery Elite) and an anthropomorphic phantom (RS-550) were used to acquire the effective dose caused by CT in PET/CT.CT acquisition conditions were as follows in both ST-16 and Elite PET/CT:tube voltage 120 kVp, pitch 1.375,rotation speed 0.8 s,noise index ranged from 8 to 30 with an interval of 2,the low limit of current was 30mA,and the high limit ranged from 200 mAto 350 mA with an interval of 50 mA. Simulating the clinical head-neck and body conditions, the phantom was scanned, and the dose-length product(DLP) was recorded to estimate the effective dose (EDCT) of CT.Results With the same acquisition conditions, the effective dose decreased along with the increasing noise index, and the curve became steeper with increasing high limit of tube current. The whole body effective dose caused by CT was lower in Elite PET/CT than that in ST-16 with the same acquisition conditions.Conclusion The effective dose caused by CT in PET/CT differs from variable scan conditions. It shows that the radiation dose can be effectively reduced by choosing appropriate acquisition conditions according to different clinical needs.
|
Received: 14 December 2017
|
|
|
|
|
[1] 臧士明,王峰,黄悦,等.68Ga-PSMA-11 PET/CT对去势抵抗性前列腺癌的诊断价值[J].中华核医学与分子影像杂志,2017,37(3):142-146. [2] 丁重阳, 李海,刘冲,等.皮下脂膜炎样T细胞淋巴瘤的18F-FDG PET/CT表现及临床分析[J]. 中国实验血液学杂志,2017,25(4):1047-52. [3] 滕才钧,邓燕云,韦建林,等.18F-FDG PET/CT诊断梗阻性黄疸的价值[J].中国医学影像技术,2017,33(12):1840-43. [4] 梁子威,耿建华,王弈斌,等.PET/CT受检者的有效剂量及其影响因素的研究[J].中国医学装备, 2015,12(11):1-5. [5] 张连宇,耿建华.PET/CT中的CT剂量和质量控制[J].中国医学影像技术, 2011,27(11):2365-2367. [6] IEC 61223-3-5-2004,Acceptance tests-Imaging performance of computed tomography X-ray equipment.2004. [7] 吴一田,耿建华.PET-CT、SPECT及SPECT-CT受检者辐射水平及其生物效应研究进展[J].中国医学装备,2017,14(3):141-146. [8] 梁子威,耿建华,王弈斌,等.PET/CT中CT仿真人体模型的图像质量与辐射剂量关系的研究[J].中华放射医学与防护杂志,2016,36(7):524-529. [9] 岳保荣,牛延涛.多排探测器计算机X线体层摄影患者剂量控制[M].北京:人民军医出版社,2011 [10] Larson DB,Wang LL,Podberesky DJ,et al.System for Verifiable CT radiation Dose Optimization Based on image Quality.Part I.Optimization Model[J].Radiology,2013,269(1):167-176. [11] Delbeke D, Coleman R E, Guiberteau MJ, et al. Procedure Guideline for Tumor Imaging with 18F-FDG PET/CT 1.0[J]. J Nucl Med, 2006,47(5):885-893. [12] 王风,赵伟,赵启超,等.核医学检查受检者所受辐射剂量分析[J].中华放射医学与防护杂志,2015,35(9):705-708. [13] 李玲,韩萍萍,续蕊,等.部分成人体部PET/CT辐射剂量评估[J].中国医疗设备,2015,30(9):104-107. [14] Quinn B,Dauer Z,Pandit-Taskar N,et al.Radiation dosimetry of 18F-FDG PET/CT:incorporating exam-specific parameters in dose estimates[J].BMC Medical Imaging,2016,16(1):41. [15] Zaman MU,Fatima N,Zaman A,et al.Significantly Low Effective Dose from 18 FDG PET/CT ScansUsing Dose Reducing Strategies:"Lesser is Better"[J].Asian Pac J Cancer Prev,2016,17(7):3465-3468. [16] 彭刚,曾勇明,罗天友.中国人仿真胸部体模检测多层螺旋CT扫描组织器官剂量的研究[J].中华放射学杂志,2011,45(6):584-588. [17] Rizzo S,Kalra M,Schmidt B,et al.Comparison of angular and combined automatic tube current modulation techniques with constant tube current CT of abdomen and pelvis[J].AJR,2006,186(3):673-679. [18] 王慧春,陈晓红,崔兰兰,等.利尿后18F-FDG PET/CT延迟显像探测男性直肠癌局部复发的临床价值[J].中华核医学与分子影像杂志,2015,35(6):446-450. [19] 韩萍萍,郑玉民,刘晓建,等.肺硬化性血管瘤CT及18F-FDG PET/CT影像特征分析[J].中华核医学与分子影像杂志,2015,35(4):251-253. [20] 许泽清,段晓蓓,樊卫.18F-FDG PET/CT显像诊断非小细胞肺癌纵隔淋巴结转移的临床价值[J].中华核医学与分子影像杂志,2013,33(5):381-382. [21] Jani AB,Schreibmann E,Rossi PJ, et al.Impact of 18F-Fluciclovine PET on Target Volume Definition for Postprostatectomy Salvage Radiotherapy:Initial Findings from a Randomized Tria[J].J Nucl Med,2017,58(3):412-418. [22] Van Der Gucht A,Aoun Sebaiti M,Guedj E,et al. Brain 18F-FDG PET Metabolic Abnormalities in Patients with Long-Lasting Macrophagic Myofascitis[J].J Nucl Med,2017,58(3):492-498. [23] Wan MYS,Endozo R,Michopoulou S,et al.PET/CT Imaging of Unstable Carotid Plaque with 68Ga-Labeled Somatostatin Receptor Ligand[J].J Nucl Med,2017,58(5):774-780. [24] Skoura E,Skoura E,Mohmaduvesh M,et al.The Impact of 68Ga-DOTATATE PET/CT Imaging on Management of Patients with Neuroendocrine Tumors:Experience from a National Referral Center in the United Kingdom[J].J Nucl Med,2016 Jan;57(1):34-40. [25] Chen CM,Lin YY,Hsu MY,et al.Performance of adaptive iterative dose reduction 3D integrated with automatic tube current modulation in radiation dose and image noise reduction compared with filtered-back projection for 80-kVp abdominal CT:Anthropomorphic phantom and patient study[J].Eur J Radiol,2016,85(9):1666-1672. [26] Zhao YX,Zuo ZW,Suo HN,et al.CT Pulmonary Angiography Using Automatic Tube Current Modulation Combination with Different Noise Index with Iterative Reconstruction Algorithm in Different Body Mass Index:Image Quality and Radiation Dose[J].Acad Radiol,2016,23(12):1513-1520. [27] 陆锐,何健,周科峰,等.自动管电流调制扫描及ASIR重建下胸部CT噪声指数的优化[J].医学影像杂志,2016,26(6):1012-1016. [28] 宋建,李仁民,李小虎,等.ASIR算法结合自动管电流调制技术在泌尿系CT平扫中的应用[J].安徽医学,2016,37(8):933-936. |
|
|
|