Analysis of performance test results of some medical linear accelerators in Hebei Province, China, 2019—2021
LI Peng1, ZHANG Yuanyuan1, JIANG Lihong1, LI Xiaopeng1, WANG Youcheng1, LIU Yong1, ZHANG Yu2, NIU Qingguo1
1. Shijiazhuang Center for Disease Control and Prevention, Shijiazhuang 050031 China; 2. Kaijiefangda Detection Technology Hebei Limited Company, Shijiazhuang 050000 China
李鹏, 张圆圆, 江丽红, 李小鹏, 王有成, 刘永, 张雨, 牛庆国. 2019—2021年河北省部分医用直线加速器性能检测结果分析[J]. 中国辐射卫生, 2022, 31(5): 592-596.
LI Peng, ZHANG Yuanyuan, JIANG Lihong, LI Xiaopeng, WANG Youcheng, LIU Yong, ZHANG Yu, NIU Qingguo. Analysis of performance test results of some medical linear accelerators in Hebei Province, China, 2019—2021. Chinese Journal of Radiological Health, 2022, 31(5): 592-596.
[1] 段杏丽, 尹俊清. 河北省医疗机构放射治疗现状调查与分析[J]. 河北医药,2016,38(23):3651-3653 Duan XL, Yin JQ. Nvestigation and analysis of radiotherapy status in medical institutions of Hebei Province[J]. Hebei Med J, 2016, 38(23): 3651-3653 [2] 安艳, 陈如松. 国内外核与放射突发事故和医学应急体系及对我们的启示[J]. 中国辐射卫生,2007,16(3):373-376. DOI: 10.13491/j.issn.1004-714X(2007)03-0373-04 An Y, Chen RS. Nuclear and radiation accidents and medical emergency system at home and abroad and its enlightenment to us[J]. Chin J Radiol Health, 2007, 16(3): 373-376. DOI: 10.13491/j.issn.1004-714X(2007)03-0373-04 [3] 祝小惠, 韩军, 梁志文, 等. 2018年湖北省放疗基本情况与质控设备配置调查分析[J]. 中国辐射卫生,2020,29(3):224-228. DOI: 10.13491/j.issn.1004-714X.2020.03.006 Zhu XH, Han J, Liang ZW, et al. Investigation and analysis on the basic situation of radiotherapy and the configuration of quality control equipment in Hubei Province in 2018[J]. Chin J Radiol Health, 2020, 29(3): 224-228. DOI: 10.13491/j.issn.1004-714X.2020.03.006 [4] 国家卫生健康委员会. WS 674—2020 医用电子直线医用直线加速器质量控制检测规范[S]. 北京: 中国标准出版社, 2020. National Health Commission. WS 674—2020 Specification for testing of quality control in medical linear accelerator[S]. Beijing: Standards Press of China, 2020. [5] 中华人民共和国国家质量监督检验检疫总局, 中国国家标准化管理委员会. GB 15213—2016 医用电子加速器性能和试验方法[S]. 北京: 中国标准出版社, 2016. General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China, Standardization Administration of the People's Republic of China. GB 15213—2016 Medical electron accelerators functional performance characteristics and test methods[S]. Beijing: Standards Press of China, 2016. [6] 徐恩慧. 医用电子直线加速器质量控制的措施[J]. 医疗装备,2018,31(6):140. DOI: 10.3969/j.issn.1002-2376.2018.06.106 Xu EH. Measures for quality control of medical electron linear accelerators[J]. Med Equip, 2018, 31(6): 140. DOI: 10.3969/j.issn.1002-2376.2018.06.106 [7] 张京战, 周开建, 赵智慧, 等. 河北省部分医用电子加速器性能检测结果与分析[J]. 中国辐射卫生,2011,20(2):209-210. DOI: 10.13491/j.cnki.issn.1004-714X.2011.02.048 Zhang JZ, Zhou KJ, Zhao ZH, et al. Analysis of quality control test results of medical electron linear accelertators in Hebei Province[J]. Chin J Radiol Health, 2011, 20(2): 209-210. DOI: 10.13491/j.cnki.issn.1004-714X.2011.02.048 [8] 王赟, 邬家龙, 张涵宇, 等. 甘肃省医用电子直线加速器质量控制及防护检测与评价[J]. 疾病预防控制通报,2020,35(5):78-80. DOI: 10.13215/j.cnki.jbyfkztb.2007016 Wang Y, Wu JL, Zhang HY, et al. Quality control and protective testing and evaluation of medical electron linear accelerator in Gansu Province[J]. Bull Dis Control Prev, 2020, 35(5): 78-80. DOI: 10.13215/j.cnki.jbyfkztb.2007016 [9] 董颖, 吴应宇, 唐孟俭, 等. 2017—2019年广西省部分医用电子直线加速器质量控制检测结果分析[J]. 中国辐射卫生,2021,30(4):407-411. DOI: 10.13491/j.issn.1004-714X.2021.04.004 Dong Y, Wu YY, Tang MJ, et al. Analysis of quality control test results of medical electron linear accelerators in Guangxi Province 2017—2019[J]. Chin J Radiol Health, 2021, 30(4): 407-411. DOI: 10.13491/j.issn.1004-714X.2021.04.004 [10] 胡波, 李长虹, 刘冉, 等. 四川省医用直线加速器质量控制检测结果[J]. 医疗装备,2019,32(7):43-44. DOI: 10.3969/j.issn.1002-2376.2019.07.023 Hu B, Li CH, Liu R, et al. Quality control test results of medical linear accelerator in Sichuan province[J]. Med Equip, 2019, 32(7): 43-44. DOI: 10.3969/j.issn.1002-2376.2019.07.023 [11] 杨迪, 李盛, 高金霞, 等. 2017—2019年兰州市放射诊疗设备质量控制现状分析[J]. 中国工业医学杂志,2021,34(5):433-435. DOI: 10.13631/j.cnki.zggyyx.2021.05.016 Yang D, Li S, Gao JX, et al. Analysis on quality control Status of radiation diagnosis and treatment equipments in Lanzhou city from 2017 to 2019[J]. Chin J Ind Med, 2021, 34(5): 433-435. DOI: 10.13631/j.cnki.zggyyx.2021.05.016 [12] 邓磊, 周宁, 邱诗宝, 等. 江西省医用电子加速器性能检测与分析[J]. 中国辐射卫生,2017,26(5):552-553. DOI: 10.13491/j.cnki.issn.1004-714X.2017.05.017 Deng L, Zhou N, Qiu SB, et al. Performance test and analysis on medical electron accelerators in Jiangxi[J]. Chin J Radiol Health, 2017, 26(5): 552-553. DOI: 10.13491/j.cnki.issn.1004-714X.2017.05.017 [13] 张京战, 周开建, 高艳辉, 等. 河北省部分医用电子加速器输出量检测结果与分析[J]. 中国辐射卫生,2016,25(5):567-568,572. DOI: 10.13491/j.cnki.issn.1004-714X.2016.05.026 Zhang JZ, Zhou KJ, Gao YH, et al. Results and analysis of partial medical electron accelerators Output test in Hebei Province[J]. Chin J Radiol Health, 2016, 25(5): 567-568,572. DOI: 10.13491/j.cnki.issn.1004-714X.2016.05.026 [14] International Atomic Energy Agency. Absorbed dose determination in photon and electron beams. An international code of practice. Technical reports series no. 277[M]. Vienna: IAEA, 1997. [15] International Atomic Energy Agency. Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water. Technical reports series no. 398[R]. Vienna: IAEA, 2000.