Influence of 4D CT-based respiratory signal acquisition methods on delineation of moving tumor targets |
LIU Qianqian, YAO Shengyu, CHEN Xuming, HOU Lingtong, HU Zhekai |
Shanghai General Hospital, Shanghai 201600 China |
|
|
Abstract Objective To compare the effects of different respiratory signal acquisition methods on the delineation of moving tumor targets. Methods A cube phantom containing a sphere was placed on a motion platform to simulate respiratory movement by setting motion period, frequency, and direction. Respiratory signal was acquired by real-time position management (RPM) method and GE method independently. Target delineation was conducted using the maximum intensity projection (MIP) sequence. The difference between the reconstructed volume and the theoretical moving volume was compared under the two respiratory signal acquisition methods for cube and sphere targets. Results Under the same respiratory signal acquisition method, the same respiratory amplitude, and different respiratory frequencies, reconstructed volume changes were relatively small. For the sphere target, the deviation between the reconstructed volume and the theoretical moving volume was −1.5% to 5.7% with the RPM method and −1.3% to −13.8% with the GE method (both P < 0.05). For the cube target, the deviation between the reconstructed volume and the theoretical moving volume was 0.2% to 0.9% with the RPM method and −2.6% to 0.9% with the GE method, with no statistical significance. Conclusion For small-volume sphere targets, the target volumes obtained from MIP images by the two respiratory signal acquisition methods are both smaller than the actual moving volume. For large-volume cube targets, there is no significant difference between the reconstructed and theoretical results with any respiratory signal acquisition method. The RPM method produces smaller deviation and better image quality when reconstructing small-volume targets.
|
Received: 13 July 2022
|
|
|
|
|
[1] Zubizarreta EH, Fidarova E, Healy B, et al. Need for radiotherapy in low and middle income countries - the silent crisis continues[J]. Clin Oncol, 2015, 27(2): 107-114. DOI: 10.1016/j.clon.2014.10.006 [2] 刘思涵, 张大昕, 宋颖辉. PPARγ与放射治疗联合应用的前景[J]. 现代生物医学进展,2012,12(29):5786-5788,5800. DOI: 10.13241/j.cnki.pmb.2012.29.046 Liu SH, Zhang DX, Song YH. The prospect of the combination use of PPARγ and radiotherapy[J]. Prog Mod Biomed, 2012, 12(29): 5786-5788,5800. DOI: 10.13241/j.cnki.pmb.2012.29.046 [3] 王银亮, 王慧涛, 郑安平, 等. TOMO-HD图像引导不同部位肿瘤放疗摆位误差的研究[J]. 中国辐射卫生,2021,30(3):331-338. DOI: 10.13491/j.issn.1004-714X.2021.03.015 Wang YL, Wang HT, Zheng AP, et al. Study on setup errors for different body carcinoma radiotherapy with image guidance in TOMO-HD[J]. Chin J Radiol Health, 2021, 30(3): 331-338. DOI: 10.13491/j.issn.1004-714X.2021.03.015 [4] 刘茹佳, 陈颖, 钟志鹏, 等. 两种算法在非小细胞肺癌调强放疗中的剂量学比较[J]. 中国辐射卫生,2019,28(6):705-708. DOI: 10.13491/j.issn.1004-714X.2019.06.029 Liu RJ, Chen Y, Zhong ZP, et al. Dosimetric comparison of two algorithms in intensity-modulated radiation therapy for NSCLC[J]. Chin J Radiol Health, 2019, 28(6): 705-708. DOI: 10.13491/j.issn.1004-714X.2019.06.029 [5] Borm KJ, Oechsner M, Wiegandt M, et al. Moving targets in 4D-CTs versus MIP and AIP: comparison of patients data to phantom data[J]. BMC Cancer, 2018, 18(1): 760. DOI: 10.1186/s12885-018-4647-4 [6] 黄玉玲, 潘纯国, 袁星星, 等. 4D-CT在肺转移瘤个体化精准放疗中的研究[J]. 实用癌症杂志,2021,36(12):2007-2010. DOI: 10.3969/j.issn.1001-5930.2021.12.026 Huang YL, Pan CG, Yuan XX, et al. Internal target volume determined with 4D-CT in lung metastases[J]. Pract J Cancer, 2021, 36(12): 2007-2010. DOI: 10.3969/j.issn.1001-5930.2021.12.026 [7] Erridge SC, Seppenwoolde Y, Muller SH, et al. Portal imaging to assess set-up errors, tumor motion and tumor shrinkage during conformal radiotherapy of non-small cell lung cancer[J]. Radiother Oncol, 2003, 66(1): 75-85. DOI: 10.1016/S0167-8140(02)00287-6 [8] 戴威, 邓小武, 黄劭敏, 等. 放疗靶区呼吸运动对CT模拟定位图像重建的几何体积精度的体模实验研究[J]. 中国医学物理学杂志,2007,24(6):391-395,414. DOI: 10.3969/j.issn.1005-202X.2007.06.001 Dai W, Deng XW, Huang SM, et al. Precision analyzing on virtual volume by using respiration motion phantom[J]. Chin J Med Phys, 2007, 24(6): 391-395,414. DOI: 10.3969/j.issn.1005-202X.2007.06.001 [9] Zhang JJ, Srivastava S, Wang CY, et al. Clinical evaluation of 4D MRI in the delineation of gross and internal tumor volumes in comparison with 4DCT[J]. J Appl Clin Med Phys, 2019, 20(9): 51-60. DOI: 10.1002/acm2.12699 [10] Dai HD, Dong LF, Lv BW, et al. Feasibility study of permanent magnet-based tumor tracking technique for precise lung cancer radiotherapy[J]. IEEE Trans Instrum Meas, 2021, 70: 4002010. DOI: 10.1109/TIM.2020.3039645 [11] Dupre PJ, Fitzherbert CM, Hand CM, et al. A new technique using 4DCT scans to create custom cardiac blocks for left-sided breast cancer treatments[J]. Pract Radiat Oncol, 2019, 9(6): e572-e578. DOI: 10.1016/j.prro.2019.06.002 [12] McWilliam A, Lee L, Harris M, et al. Benefit of using motion compensated reconstructions for reducing inter-observer and intra-observer contouring variation for organs at risk in lung cancer patients[J]. Radiother Oncol, 2018, 126(2): 333-338. DOI: 10.1016/j.radonc.2017.11.021 [13] Koksal C, Kesen ND, Okutan M, et al. Investigation of approaches for internal target volume definition using 4-dimensional computed tomography in stereotactic body radiotherapy of lung cancer[J]. Med Dosim, 2021, 46(2): 136-142. DOI: 10.1016/j.meddos.2020.10.004 [14] Li G, Wei J, Li D, et al. SU-E-J-120: Characterization of 4D lung ventilation based on 4DCT lung density[J]. Med Phys, 2019, 39(6Part7): 3680. DOI: 10.1118/1.4734956 [15] Mizuno K, Song YL, Obcemea C. Comparison of the sensitivities of two 4DCT techniques: RPM and bellows system[J]. Int J Biomed Eng Technol, 2012, 8(2/3): 220-229. DOI: 10.1504/IJBET.2012.046087 [16] Chung H, Jung J, Jeong C, et al. Evaluation of delivered dose to a moving target by 4D dose reconstruction in gated volumetric modulated arc therapy[J]. PLoS One, 2018, 13(9): e0202765. DOI: 10.1371/journal.pone.0202765 [17] Rabe M, Thieke C, Düsberg M, et al. Real-time 4DMRI-based internal target volume definition for moving lung tumors[J]. Med Phys, 2020, 47(4): 1431-1442. DOI: 10.1002/mp.14023 [18] Chi YW, Liang J, Qin X, et al. Respiratory motion sampling in 4DCT reconstruction for radiotherapy[J]. Med Phys, 2012, 39(4): 1696-1703. DOI: 10.1118/1.3691174
|
|
|
|