Study on Relationship between Mass and ESR Signal Intensity of Enamel |
JIAO Ling1, ZHANG Wen-yi1, PAN Zhi-hong2, et al |
Institute of Radiation Medicine, Chinese Academy of Medical Sciences and Peking Union Medical College, Tianjin 300192, China |
|
|
Abstract Objective To study the relationship of different enamel sample mass and their enamel ESR signal intensities in order to find out a suitable mass range used in dose reconstruction.Methods The enamel samples are prepared from non-irradiated adult teeth by mechanical crash.The samples with different mass are scanned by an ESR spectrometer after their irradiation by 60Co γ rays.Results The linearity between ESR signal response and mass is not good in the range of 10~200 mg.But the linearity between ESR signal response and mass is good in the range of 40~120 mg.Conclusion It is better to choose enamel samples with the mass in the range of 40~120 mg in dose reconstruction.
|
Received: 08 November 2004
|
|
|
|
|
|
|
|