摘要目的 估算双源CT两种扫描模式下的有效剂量。方法 在仿真人体模内布放热释光剂量计,分别于两种模式下扫描体模,得到双源CT两种扫描模式下的组织、器官的吸收剂量,并根据国际放射防护委员会(International Commission on Radiological Protection ICRP)103号出版物中推荐的组织权重因子估算受检者的有效剂量。同时采用美国医学物理师协会(American Association of Physicists in Medicine AAPM)推荐的剂量长度乘积(Dose Length Product DLP)乘以K值法估算有效剂量。比较两种方法得到的有效剂量大小。结果 得到两种扫描模式下的有效剂量。两种扫描模式扫描体模估算有效剂量分别为3.66和2.98 mSv。经由AAPM推荐K值计算有效剂量两种扫描模式分别为2.81和1.55 mSv,均低于经体模扫描估算的有效剂量。结论 双源CT头部扫描时眼晶状体吸收剂量较高。AAPM推荐的DLP乘以K值法估算的有效剂量较仿真人法低。
[1] ICRP.Managing patient dose in multi-detector computed tomography(MDCT)[R].ICRP Publication 102.Oxford:Pergamon Press,2007:34. [2] 李敏,郑理华,张晨,等.双源CT图像质量与辐射剂量的关系与分析[J].医疗卫生装备,2013,34(12):88-89. [3] 李剑,宦怡,印弘,等.低浓度对比剂和超低辐射剂量在双源CT冠状动脉成像中的应用[J].实用放射学杂志,2015,31(2):300-304. [4] ICRP.The 2007 Recommendations of the International Commission on Radiological Protection[R].ICRP Publication 103.2007:63-70. [5] DeMarco JJ,Cagnon CH,Cody DD,et a1.Estimating radiation doses from mulltidetector CT using Monte Carlo simulations:effects of different size voxelized patient models on magnitudes of organ and effective dose[J].Phys Med Biol,2007,52(9):2583-2597. [6] 张琳,朱建国,闵楠,等.3种常见介入诊疗中放射工作人员有效剂量的估算[J].中华放射医学与防护杂志,2011,,3(4):393. [7] 李士俊.电离辐射剂量学基础[M].苏州:苏州大学出版社,2008:43. [8] McCollough C,Cody D,et al.The Measurement,Reporting,and Management of Radiation Dose in CT[R]:Report of AAPM Task Group 23 of the Diagnostic Imaging Council CT Committee.USA:AAPM.2008:13. [9] 李士俊.电离辐射剂量学基础[M].苏州:苏州大学出版社,2008:31. [10] Alkadhi H,Stolzmann P,Desbiolle L,et al.Low-dose,128-slice,dual-source CT coronary angiography:accuracy andradiation dose of the high-pitch and the step-and-shoot mode[J].Heart.2010,96(12):933-938. [11] Han BK,Grant KL,Garberich R,et al.Assesment of an iterative reconstruction algorithm(SAFIRE) on image quality in pediatric cardiac CT datasets[J].J Cardiovasc Comput Tomogr,2012,6(3):200-204. [12] DeMarco JJ,Cagnon CH,Cody DD,et a1.Estimating radiation doses from mulltidetector CT using Monte Carlo simulations:effects of different size voxelized patient models on magnitudes of organ and effective dose[J].Phys Med Biol,2007,52(9):2583-2597. [13] Wall BF.Radiation protection dosimetry for diagnostic radiology patients[J].Radiat Prot Dosim,2004,109(4):409-419. [14] Zelikman M, Kruchinin S, Burashov V. How to Estimate Multislice CT Effective Dose More Correctly[J]. Health Phys,2013,105(6):534-539.